Can the radiopaque marker in surgical swabs scratch orthopaedic implant surfaces?
Implant
Metal
Radiopaque marker
Swab
Titanium
Wear
X-ray
Journal
European journal of orthopaedic surgery & traumatology : orthopedie traumatologie
ISSN: 1432-1068
Titre abrégé: Eur J Orthop Surg Traumatol
Pays: France
ID NLM: 9518037
Informations de publication
Date de publication:
Feb 2019
Feb 2019
Historique:
received:
31
05
2018
accepted:
12
09
2018
pubmed:
1
10
2018
medline:
31
5
2019
entrez:
1
10
2018
Statut:
ppublish
Résumé
To determine whether the radiopaque marker strip, which is woven in surgical swabs, causes measureable wear on metal implants at pressures typically used to wipe off fluid from their surface. Finger pressure used to wipe a surface was measured and used as a reference pressure for further testing. A tribological wear rig was then used to analyse the wear caused on polished titanium plates by a cobalt chromium pin (the control test), the pin covered by a surgical swab and the pin covered by a radiopaque marker strip. It was found that the cotton part or the radiopaque marker of surgical swabs on polished medical grade titanium plates caused no significant wear. In contrast severe scratching was observed from the cobalt chromium pin on its own. To our knowledge, this is the first study in the literature analysing the wear caused by the surgical swabs and radiopaque strip on metal implants. The results suggest that surgical swabs are safe to use on metallic implants at pressures typical of a wiping motion.
Identifiants
pubmed: 30269182
doi: 10.1007/s00590-018-2309-5
pii: 10.1007/s00590-018-2309-5
doi:
Substances chimiques
Titanium
D1JT611TNE
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
383-388Références
World J Surg. 2000 May;24(5):521-7
pubmed: 10787070
J Bone Joint Surg Br. 2001 Aug;83(6):902-11
pubmed: 11521937
Proc Inst Mech Eng H. 2001;215(6):605-10
pubmed: 11848393
J Arthroplasty. 2004 Jan;19(1):68-77
pubmed: 14716654
Biomaterials. 2004 Aug;25(17):3511-22
pubmed: 15020125
Proc Inst Mech Eng H. 2006 Feb;220(2):385-98
pubmed: 16669404
Arch Orthop Trauma Surg. 2008 Oct;128(10):1167-8
pubmed: 17906867
J Appl Biomater Biomech. 2010 Jan-Apr;8(1):1-6
pubmed: 20740415
Am J Clin Pathol. 2010 Dec;134(6):886-93
pubmed: 21088151
Clin Cases Miner Bone Metab. 2013 Jan;10(1):34-40
pubmed: 23858309
Arch Bone Jt Surg. 2013 Sep;1(1):41-3
pubmed: 25207284
Clin Orthop Relat Res. 1995 Feb;(311):46-53
pubmed: 7634590
Clin Orthop Relat Res. 1993 Sep;(294):361-78
pubmed: 8358943