A randomised controlled trial to compare two coronary pressure wires using simultaneous measurements in human coronary arteries: the COMET trial.
Journal
EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology
ISSN: 1969-6213
Titre abrégé: EuroIntervention
Pays: France
ID NLM: 101251040
Informations de publication
Date de publication:
08 Feb 2019
08 Feb 2019
Historique:
pubmed:
31
10
2018
medline:
31
7
2019
entrez:
31
10
2018
Statut:
epublish
Résumé
We aimed to examine the relative performance of the new COMET wire from Boston Scientific (BS), and the established technology from St. Jude/Abbott Vascular (SJ). We compared simultaneous readings from pairs of wires. Patients were randomised to one of three groups: BS/BS, SJ/SJ, or SJ/BS. The last group was sub-randomised to specify the type of wire that would be passed first. After pressure equalisation at the guide catheter, we recorded paired observations in sequence: (a) distal to proximal pressure ratio at baseline, (b) FFR at maximum hyperaemia, and (c) pressure on withdrawal into the guide catheter to quantify "drift". We randomised 106 patients, yielding 288 sets of paired recordings (BS/BS=90; SJ/SJ=90; SJ/BS=108). Drift was recorded from 208 vessels (BS=105; SJ=103). All wires were successfully advanced to their desired positions in the coronary vasculature. The mean (±SD) differences for the randomised pairs were similar: BS/BS=0.0016 (0.023); SJ/SJ=0.002 (0.03); SJ/BS=0.0013 (0.028). The primary outcome tested the hypothesis that the absolute magnitude of the difference (irrespective of sign) observed in the SJ/BS pairing would be similar to that in the SJ/SJ group. The median (IQR) values were SJ/BS=0.015 (0.01-0.03); SJ/SJ=0.01 (0.00-0.03); p=0.61. The drift, expressed as the median (IQR) difference in Pd/Pa from 1.0 (irrespective of sign), was similar: BS=0.02 (0.01-0.05); SJ=0.02 (0.01-0.04); p=0.14. We found no significant difference between these wires in terms of safety and performance.
Identifiants
pubmed: 30375339
pii: EIJ-D-18-00786
doi: 10.4244/EIJ-D-18-00786
pii:
doi:
Types de publication
Journal Article
Randomized Controlled Trial
Langues