Kinematic alignment technique for medial OXFORD UKA: An in-silico study.


Journal

Orthopaedics & traumatology, surgery & research : OTSR
ISSN: 1877-0568
Titre abrégé: Orthop Traumatol Surg Res
Pays: France
ID NLM: 101494830

Informations de publication

Date de publication:
02 2019
Historique:
received: 22 09 2018
revised: 20 11 2018
accepted: 20 11 2018
pubmed: 1 1 2019
medline: 14 1 2020
entrez: 1 1 2019
Statut: ppublish

Résumé

Mobile bearing unicompartmental knee arthroplasty (UKA) Oxford™ components are recommended to be systematically and mechanically aligned (MA) for restoring the constitutional lower-limb alignment. Good long-term clinical outcomes have been generated with the medially implanted MA Oxford™, but some sub-optimal biomechanical-related complications still remain. Kinematic Alignment (KA) is a personalised technique for anatomically and kinematically implanting components (total knee, fixed bearing partial knee, total hip) aimed at creating more physiological prosthetic joint biomechanics. Interestingly, for decades the principles for implanting fixed bearing UKA components were consistent with those promoted by the KA technique, but differently formulated. We initiated this computational study to assess the feasibility of this technique with the Oxford™ components, as we thought this more anatomical implantation may be clinically advantageous. We surmised that kinematically aligning the Oxford™ medial UKA would maximise the prosthesis-bone interface through maximising the implants' size used (question 1), and alter, within an acceptable limit, the components' orientation (question 2) compared to conventional mechanical alignment. A cohort of 40 consecutive medial osteoarthritic knee patients scheduled for UKA had a preoperative CT scan that was segmented to create 3D knee bone models. MA and KA of medial UKA Oxford We found no difference in component size, but significantly fewer occurrences of borderline fit with the KA simulation. KA technique oriented the femoral component 3.6° more valgus (from 1° varus to 7° valgus) and the tibial component 2.9° more varus (from 8° varus to 0°) compared to the MA technique. The tibial component slope in KA simulation was 6.4° posterior (from 0 to 12°) compared to a systematic 7° posterior for MA positioning. Kinematic alignment of the medial Oxford™ generated a different, albeit still acceptable (Oxford group recommendations), implant orientation, in addition to a likely better shape-fit between components and the supportive bone cut, compared to the MA technique. The potential to improve the implants' interaction and to restore a more physiological bone loading makes the KA of Oxford™ an attractive, potentially clinically beneficial option. Clinical investigations are needed to assess its true value. I, computational study.

Sections du résumé

BACKGROUND
Mobile bearing unicompartmental knee arthroplasty (UKA) Oxford™ components are recommended to be systematically and mechanically aligned (MA) for restoring the constitutional lower-limb alignment. Good long-term clinical outcomes have been generated with the medially implanted MA Oxford™, but some sub-optimal biomechanical-related complications still remain. Kinematic Alignment (KA) is a personalised technique for anatomically and kinematically implanting components (total knee, fixed bearing partial knee, total hip) aimed at creating more physiological prosthetic joint biomechanics. Interestingly, for decades the principles for implanting fixed bearing UKA components were consistent with those promoted by the KA technique, but differently formulated. We initiated this computational study to assess the feasibility of this technique with the Oxford™ components, as we thought this more anatomical implantation may be clinically advantageous.
HYPOTHESIS
We surmised that kinematically aligning the Oxford™ medial UKA would maximise the prosthesis-bone interface through maximising the implants' size used (question 1), and alter, within an acceptable limit, the components' orientation (question 2) compared to conventional mechanical alignment.
METHODS
A cohort of 40 consecutive medial osteoarthritic knee patients scheduled for UKA had a preoperative CT scan that was segmented to create 3D knee bone models. MA and KA of medial UKA Oxford
RESULTS
We found no difference in component size, but significantly fewer occurrences of borderline fit with the KA simulation. KA technique oriented the femoral component 3.6° more valgus (from 1° varus to 7° valgus) and the tibial component 2.9° more varus (from 8° varus to 0°) compared to the MA technique. The tibial component slope in KA simulation was 6.4° posterior (from 0 to 12°) compared to a systematic 7° posterior for MA positioning.
DISCUSSION AND CONCLUSION
Kinematic alignment of the medial Oxford™ generated a different, albeit still acceptable (Oxford group recommendations), implant orientation, in addition to a likely better shape-fit between components and the supportive bone cut, compared to the MA technique. The potential to improve the implants' interaction and to restore a more physiological bone loading makes the KA of Oxford™ an attractive, potentially clinically beneficial option. Clinical investigations are needed to assess its true value.
LEVEL OF EVIDENCE
I, computational study.

Identifiants

pubmed: 30595413
pii: S1877-0568(18)30361-X
doi: 10.1016/j.otsr.2018.11.005
pii:
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

63-70

Informations de copyright

Copyright © 2018 Elsevier Masson SAS. All rights reserved.

Auteurs

Charles Rivière (C)

The MSK Lab-Imperial college London, South West London Elective Orthopaedic Centre, London, United Kingdom. Electronic address: c.riviere@imperial.ac.uk.

Ciara Harman (C)

South West London Elective Orthopaedic Centre, Dorking road, KT18 7EG Epsom, United Kingdom.

Anthony Leong (A)

The MSK Lab-Imperial college London, Charing Cross Campus, Laboratory Block, W6 8RP London, United Kingdom.

Justin Cobb (J)

The MSK Lab-Imperial college London, Charing Cross Campus, Laboratory Block, W6 8RP London, United Kingdom.

Cedric Maillot (C)

The MSK Lab-Imperial college London, South West London Elective Orthopaedic Centre, London, United Kingdom.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH