CHARACTERIZATION OF A LITHIUM FORMATE EPR-DOSIMETRY SYSTEM FOR PROTON RADIATION THERAPY.
Journal
Radiation protection dosimetry
ISSN: 1742-3406
Titre abrégé: Radiat Prot Dosimetry
Pays: England
ID NLM: 8109958
Informations de publication
Date de publication:
31 Dec 2019
31 Dec 2019
Historique:
received:
25
08
2018
revised:
25
11
2018
pubmed:
10
1
2019
medline:
5
8
2020
entrez:
10
1
2019
Statut:
ppublish
Résumé
The specific aim for the characterization of the lithium formate dosimetry system is to determine response and stability in a proton beam. The long-term goal for this investigation is an audit system for proton therapy like the end-to-end dose determinations performed for radiotherapy with photons. For a 150-MeV proton beam, the dose response was found to be linear in the dose interval 0-8.8 Gy. The accuracy of dose reconstruction was controlled in a blind test, in which the dose of 6.63 Gy was measured in samples irradiated with a real dose of 6.70 Gy. The stability was determined by irradiations of sets of four dosimeters every week during 1 month and analyzed at the same day thereafter. The fitting of the fading curve was done with a second-order polynomial resulting in a 6.6% lower value compared to the reference after 31 d.
Identifiants
pubmed: 30624734
pii: 5281159
doi: 10.1093/rpd/ncy293
doi:
Substances chimiques
Formates
0
formic acid
0YIW783RG1
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
83-87Informations de copyright
© The Author(s) 2019. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.