Cellular resolution corneal imaging with extended imaging range.
Journal
Optics express
ISSN: 1094-4087
Titre abrégé: Opt Express
Pays: United States
ID NLM: 101137103
Informations de publication
Date de publication:
21 Jan 2019
21 Jan 2019
Historique:
entrez:
31
1
2019
pubmed:
31
1
2019
medline:
20
2
2019
Statut:
ppublish
Résumé
Current optical coherence tomography (OCT) technology, which is used for imaging the eye's anterior segment, has been established as a clinical gold standard for the diagnosis of corneal diseases. However, the cellular resolution level information that is critical for many clinical applications is still not available. The major technical challenges toward cellular resolution OCT imaging are the limited ranging depth and depth of focus (DOF). In this work, we present a novel ultrahigh resolution OCT system that achieves an isotropic spatial resolution of <2 µm in tissue. The proposed system could approximately double the ranging depth and extend the DOF using the dual-spectrometer design and the forward-model based digital refocusing method, respectively. We demonstrate that the novel system is capable of visualizing the full thickness of the pig cornea over the ranging depth of 3.5 mm and the border of the corneal endothelial cells 8 times Rayleigh range away from the focal plane. This technology has the potential to realize cellular resolution corneal imaging in vivo.
Identifiants
pubmed: 30696198
pii: 404017
doi: 10.1364/OE.27.001298
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM