EP15A3 Based Precision and Trueness Verification of VITROS HbA1C Immunoassay.
EP 15 A3
HbA1c
Precision
Trueness
Verification
Journal
Indian journal of clinical biochemistry : IJCB
ISSN: 0970-1915
Titre abrégé: Indian J Clin Biochem
Pays: India
ID NLM: 8708303
Informations de publication
Date de publication:
Jan 2019
Jan 2019
Historique:
received:
02
09
2017
accepted:
06
11
2017
entrez:
8
2
2019
pubmed:
8
2
2019
medline:
8
2
2019
Statut:
ppublish
Résumé
Verification of analytical performance of measurands becomes an essential requirement for the laboratories before proceeding to patients' samples testing. In our study we have verified the performance of HbA1C Immunoturbidimetric assay (VITROS 5600) against manufacturers' claims using CLSI EP15A3 Guidelines. We performed our study using two concentrations of Quality Control from Bio-Rad (Level 1 and Level 2). A precision verification study was carried out using five replicates of QC per day for five days following which imprecision estimates in form of Within Run (Repeatability) %CV and Within Lab %CV were calculated and compared against manufacturer's claims. Second part of our study included derivation of grand mean from the results of 25 replicates of QC used for precision verification. This was compared against the Target Value of the assigned QC obtained from the peer group mean of laboratories participating in interlaboratory QC program (unityTM Interlab-Bio-Rad) for %bias estimation. The findings of our precision study showed an acceptable Within Lab imprecision (%CV
Identifiants
pubmed: 30728678
doi: 10.1007/s12291-017-0712-z
pii: 712
pmc: PMC6346623
doi:
Types de publication
Journal Article
Langues
eng
Pagination
89-94Déclaration de conflit d'intérêts
None of the authors of this study have any conflict of interest.This article contains study only with quality control materials commercially obtained. No patient samples were involved.
Références
Clin Biochem Rev. 2008 Aug;29 Suppl 1:S23-6
pubmed: 18852851