Locking Hole Inserts: Effect of Insertion Torque on Fatigue Performance and Insert Loosening in Locking Plates.


Journal

Journal of orthopaedic trauma
ISSN: 1531-2291
Titre abrégé: J Orthop Trauma
Pays: United States
ID NLM: 8807705

Informations de publication

Date de publication:
03 2019
Historique:
entrez: 16 2 2019
pubmed: 16 2 2019
medline: 26 5 2020
Statut: ppublish

Résumé

To evaluate the effect of locking hole inserts (LHIs) and their insertion torque on locking plate fatigue life. Eighteen standard 3.5-mm locking plates were instrumented with LHIs (Smith & Nephew, Memphis, TN) of 1.70 or 3.96 Nm insertion torque, or without LHIs, whereas eleven 4.5-mm locking plates were instrumented with LHIs at 3.96 Nm insertion torque or without LHIs. Plates were cyclically loaded to failure (ie, plate fracture) in four-point bending. Number of cycles to plate failure were measured. The 3.5-mm plates with 1.70 Nm LHI insertion torque had a 52% increase in cycles to failure compared with plates without LHIs (114,300 ± 23,680 vs. 75,487 ± 15,746 cycles; P = 0.01). Increasing insertion torque to 3.96 Nm led to a further increase of 36% in fatigue life (155,177 ± 32,493 cycles; P = 0.02) and a 106% increase compared with plates without LHIs (P = 0.001). The 4.5-mm plates with 3.96 Nm insertion torque had a 48% increase in cycles to failure when compared with plates without LHIs (74,369 ± 10,181 vs. 50,214 ± 5544 cycles; P = 0.001). LHIs significantly extend plate fatigue length, which would be advantageous in the setting of delayed fracture healing. We recommend the use of LHIs in round locking holes over bony gaps whenever possible; however, we recognize that these findings are limited to implants manufactured by Smith & Nephew.

Identifiants

pubmed: 30768531
doi: 10.1097/BOT.0000000000001387
pii: 00005131-201903000-00003
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

120-124

Auteurs

Joanne Y Zhang (JY)

Boston University School of Medicine, Boston, MA.

Paul Tornetta (P)

Boston University School of Medicine, Boston, MA.

Bob Jones (B)

Smith & Nephew, Memphis, TN.

Yanming Zheng (Y)

Smith & Nephew, Memphis, TN.

Andy Whitten (A)

Smith & Nephew, Memphis, TN.

Jacob Cartner (J)

Smith & Nephew, Memphis, TN.

William M Ricci (WM)

Washington University St. Louis School of Medicine, St. Louis, MO.

Articles similaires

Failed radial head arthroplasty treated by removal of the implant.

Juan Ameztoy Gallego, Blanca Diez Sanchez, Afonso Vaquero-Picado et al.
1.00
Humans Male Female Middle Aged Range of Motion, Articular
Humans Arthroplasty, Replacement, Shoulder Male Female Retrospective Studies

Hemiarthroplasty in young patients.

Hazimah Mahmud, Dong Wang, Andra Topan-Rat et al.
1.00
Humans Male Hemiarthroplasty Middle Aged Aged

Classifications MeSH