72-Hour in vivo evaluation of nitric oxide generating artificial lung gas exchange fibers in sheep.


Journal

Acta biomaterialia
ISSN: 1878-7568
Titre abrégé: Acta Biomater
Pays: England
ID NLM: 101233144

Informations de publication

Date de publication:
05 2019
Historique:
received: 03 08 2018
revised: 26 03 2019
accepted: 02 04 2019
pubmed: 7 4 2019
medline: 13 5 2020
entrez: 7 4 2019
Statut: ppublish

Résumé

The large, densely packed artificial surface area of artificial lungs results in rapid clotting and device failure. Surface generated nitric oxide (NO) can be used to reduce platelet activation and coagulation on gas exchange fibers, while not inducing patient bleeding due to its short half-life in blood. To generate NO, artificial lungs can be manufactured with PDMS hollow fibers embedded with copper nanoparticles (Cu NP) and supplied with an infusion of the NO donor S-nitroso-N-acetyl-penicillamine (SNAP). The SNAP reacts with Cu NP to generate NO. This study investigates clot formation and gas exchange performance of artificial lungs with either NO-generating Cu-PDMS or standard polymethylpentene (PMP) fibers. One miniature artificial lung (MAL) made with 10 wt% Cu-PDMS hollow fibers and one PMP control MAL were attached to sheep in parallel in a veno-venous extracorporeal membrane oxygenation circuit (n = 8). Blood flow through each device was set at 300 mL/min, and each device received a SNAP infusion of 0.12 μmol/min. The ACT was between 110 and 180 s in all cases. Blood flow resistance was calculated as a measure of clot formation on the fiber bundle. Gas exchange experiments comparing the two groups were conducted every 24 h at blood flow rates of 300 and 600 mL/min. Devices were removed once the resistance reached 3x baseline (failure) or following 72 h. All devices were imaged using scanning electron microscopy (SEM) at the inlet, outlet, and middle of the fiber bundle. The Cu-PDMS NO generating MALs had a significantly smaller increase in resistance compared to the control devices. Resistance rose from 26 ± 8 and 23 ± 5 in the control and Cu-PDMS devices, respectively, to 35 ± 8 mmHg/(mL/min) and 72 ± 23 mmHg/(mL/min) at the end of each experiment. The resistance and SEM imaging of fiber surfaces demonstrate lower clot formation on Cu-PDMS fibers. Although not statistically significant, oxygen transfer for the Cu-PDMS MALs was 13.3% less than the control at 600 mL/min blood flow rate. Future in vivo studies with larger Cu-PDMS devices are needed to define gas exchange capabilities and anticoagulant activity over a long-term study at clinically relevant ACTs. STATEMENT OF SIGNIFICANCE: In artificial lungs, the large, densely-packed blood contacting surface area of the hollow fiber bundle is critical for gas exchange but also creates rapid, surface-generated clot requiring significant anticoagulation. Monitoring of anticoagulation, thrombosis, and resultant complications has kept permanent respiratory support from becoming a clinical reality. In this study, we use a hollow fiber material that generates nitric oxide (NO) to prevent platelet activation at the blood contacting surface. This material is tested in vivo in a miniature artificial lung and compared against the clinical standard. Results indicated significantly reduced clot formation. Surface-focused anticoagulation like this should reduce complication rates and allow for permanent respiratory support by extending the functional lifespan of artificial lungs and can further be applied to other medical devices.

Identifiants

pubmed: 30953800
pii: S1742-7061(19)30238-7
doi: 10.1016/j.actbio.2019.04.004
pmc: PMC6513705
mid: NIHMS1526700
pii:
doi:

Substances chimiques

Dimethylpolysiloxanes 0
Nylons 0
poly(dimethylsiloxane)-polyamide copolymer 0
Nitric Oxide 31C4KY9ESH
Copper 789U1901C5
S-Nitroso-N-Acetylpenicillamine 79032-48-7

Types de publication

Evaluation Study Journal Article Research Support, N.I.H., Extramural

Langues

eng

Sous-ensembles de citation

IM

Pagination

122-131

Subventions

Organisme : NHLBI NIH HHS
ID : R01 HL089043
Pays : United States

Informations de copyright

Copyright © 2019 Acta Materialia Inc. Published by Elsevier Ltd. All rights reserved.

Références

Am J Clin Nutr. 1998 May;67(5 Suppl):1069S-1073S
pubmed: 9587154
Pediatr Crit Care Med. 2015 Feb;16(2):167-74
pubmed: 25647124
Vet Med Int. 2014;2014:972074
pubmed: 24895542
Acta Biomater. 2016 Jun;37:111-9
pubmed: 27095484
Biol Trace Elem Res. 2002 Jul;88(1):87-90
pubmed: 12117268
Biomaterials. 2011 Sep;32(26):5957-69
pubmed: 21696821
Biomaterials. 2007 Dec;28(34):5144-7
pubmed: 17689608
ASAIO J. 2011 Nov-Dec;57(6):539-44
pubmed: 22036723
J Thorac Cardiovasc Surg. 2006 Mar;131(3):719-23
pubmed: 16515929
Crit Care Med. 2015 Mar;43(3):654-64
pubmed: 25565460
Ann Thorac Surg. 2007 Sep;84(3):988-94
pubmed: 17720415
ASAIO J. 2005 Jul-Aug;51(4):404-11
pubmed: 16156307
ASAIO J. 2009 May-Jun;55(3):304-6
pubmed: 19282751
Biomaterials. 2005 May;26(14):1685-93
pubmed: 15576142
Org Lett. 2008 Oct 16;10(20):4593-6
pubmed: 18811172
Ann Thorac Surg. 2016 Feb;101(2):682-9
pubmed: 26443879
J Vasc Surg. 2004 Oct;40(4):803-11
pubmed: 15472611
Toxicol Lett. 2006 May 25;163(2):109-20
pubmed: 16289865
Trans Am Soc Artif Intern Organs. 1985;31:628-33
pubmed: 3837521
Biomaterials. 2013 Sep;34(28):6957-66
pubmed: 23777908
Biomaterials. 2010 Apr;31(10):2736-45
pubmed: 20042236
J Biomed Mater Res A. 2013 Dec;101(12):3511-9
pubmed: 23613156
Acta Biomater. 2016 Aug;40:153-161
pubmed: 26969525
Acta Biomater. 2016 Oct 15;44:304-12
pubmed: 27506125
Perfusion. 2007 May;22(3):193-200
pubmed: 18018399
Am J Obstet Gynecol. 1996 Jul;175(1):222-8
pubmed: 8694056
Am J Transplant. 2009 Apr;9(4):853-7
pubmed: 19344471
J Mater Chem B. 2014 Feb 28;2(8):1059-1067
pubmed: 24634777
Ann Thorac Surg. 2014 Jul;98(1):97-102
pubmed: 24856794
Med Eng Phys. 1997 Oct;19(7):618-29
pubmed: 9457695
ASAIO J. 2017 Sep/Oct;63(5):644-649
pubmed: 28719441
Ann Thorac Surg. 2011 Jan;91(1):250-5
pubmed: 21172523
Biochem J. 1978 Jan 15;170(1):137-43
pubmed: 629772
J Clin Pathol. 1968 Jan;21(1):110-2
pubmed: 5697326
ASAIO J. 1996 Sep-Oct;42(5):M604-9
pubmed: 8944952
Can Vet J. 2003 Nov;44(11):898-906
pubmed: 14664352
Am J Physiol. 1998 Jun;274(6):H2163-76
pubmed: 9841542
J Extra Corpor Technol. 2000 Dec;32(4):214-9
pubmed: 11194058
J Biomed Mater Res A. 2013 Nov;101(11):3201-10
pubmed: 23554300
Perfusion. 2013 Mar;28(2):132-7
pubmed: 23201818
Ann Thorac Surg. 2007 Oct;84(4):1136-43; discussion 1143
pubmed: 17888959

Auteurs

Angela Lai (A)

Carnegie Mellon University, United States. Electronic address: angelala@andrew.cmu.edu.

Caitlin T Demarest (CT)

Carnegie Mellon University, United States.

Chi Chi Do-Nguyen (CC)

Carnegie Mellon University, United States.

Rei Ukita (R)

Carnegie Mellon University, United States.

David J Skoog (DJ)

Carnegie Mellon University, United States.

Neil M Carleton (NM)

Carnegie Mellon University, United States.

Kagya A Amoako (KA)

University of New Haven, United States.

Patrick J Montoya (PJ)

MedArray, Inc, United States.

Keith E Cook (KE)

Carnegie Mellon University, United States.

Articles similaires

Humans Meals Time Factors Female Adult

Vancomycin-associated DRESS demonstrates delay in AST abnormalities.

Ahmed Hussein, Kateri L Schoettinger, Jourdan Hydol-Smith et al.
1.00
Humans Drug Hypersensitivity Syndrome Vancomycin Female Male
Robotic Surgical Procedures Animals Humans Telemedicine Models, Animal

Odour generalisation and detection dog training.

Lyn Caldicott, Thomas W Pike, Helen E Zulch et al.
1.00
Animals Odorants Dogs Generalization, Psychological Smell

Classifications MeSH