The potential of photon-counting CT for quantitative contrast-enhanced imaging in radiotherapy.


Journal

Physics in medicine and biology
ISSN: 1361-6560
Titre abrégé: Phys Med Biol
Pays: England
ID NLM: 0401220

Informations de publication

Date de publication:
31 05 2019
Historique:
pubmed: 19 4 2019
medline: 27 2 2020
entrez: 19 4 2019
Statut: epublish

Résumé

The aim of this study is to use a simulation environment to evaluate the potential of using photon-counting CT (PCCT) against dual-energy CT (DECT) in the context of quantitative contrast-enhanced CT for radiotherapy. An adaptation of Bayesian eigentissue decomposition by Lalonde et al (2017 Med. Phys. 44 5293-302) that incorporates the estimation of contrast agent fractions and virtual non-contrast (VNC) parameters is proposed, and its performance is validated against conventional maximum likelihood material decomposition methods for single and multiple contrast agents. PCCT and DECT are compared using two simulation frameworks: one including ideal CT numbers with image-based Gaussian noise and another defined as a virtual patient with projection-based Poisson noise and beam hardening artifacts, with both scenarios considering spectral distortion for PCCT. The modalities are compared for their accuracy in estimating four key physical parameters: (1) the contrast agent fraction, as well as VNC parameters relevant to radiotherapy such as the (2) electron density, (3) proton stopping power and (4) photon linear attenuation coefficient. Considering both simulation frameworks, a reduction of root mean square (RMS) errors with PCCT is noted for all physical parameters evaluated, with the exception of the error on the contrast agent fraction being about constant through modalities in the virtual patient. Notably, for the virtual patient, RMS errors on VNC electron density and stopping power are respectively reduced from 2.0% to 1.4% and 2.7% to 1.4% when going from DECT to PCCT with four energy bins. The increase in accuracy is comparable to the differences between contrast-enhanced and non-contrast DECT. This study suggests that in a realistic simulation environment, the overall accuracy of radiotherapy-related parameters can be increased when using PCCT with four energy bins instead of DECT. This confirms the potential of PCCT to provide robust and quantitative tissue parameters for contrast-enhanced CT required in radiotherapy applications.

Identifiants

pubmed: 30999288
doi: 10.1088/1361-6560/ab1af1
doi:

Substances chimiques

Contrast Media 0

Types de publication

Journal Article Research Support, Non-U.S. Gov't

Langues

eng

Sous-ensembles de citation

IM

Pagination

115020

Auteurs

Mikaël Simard (M)

Département de physique, Université de Montréal, Pavillon Roger-Gaudry (D-428), 2900 boulevard Édouard-Montpetit, Montréal, Québec H3T 1J4, Canada. Author to whom any correspondence should be addressed.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH