Compare the accuracy and precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the international reference flow cytometric method in platelet counting.


Journal

PloS one
ISSN: 1932-6203
Titre abrégé: PLoS One
Pays: United States
ID NLM: 101285081

Informations de publication

Date de publication:
2019
Historique:
received: 18 11 2018
accepted: 08 05 2019
entrez: 25 5 2019
pubmed: 28 5 2019
medline: 30 1 2020
Statut: epublish

Résumé

The aim of this study is to evaluate the performance of different platelet counting methods (optical, impedance, fluorescence and hand counting) applied in different analysers by comparing with the international flow cytometric reference method (IRM). A total of 333 blood samples from different subgroups (168 cases with thrombocytopenia, 136 cases with normal platelet counts and 29 cases with thrombocytosis) were tested. Regarding IRM as the gold standard, we compared the accuracy and precision of different platelet count methods; i.e. LH780 (impedance), BC-6000 Plus (optical (O) and impedance (I)), Sysmex XN-9000 (optical (O), impedance (I), fluorescence (F)), and hand counting. Sysmex XN-9000-F (r = 0.988) had the best correlation with IRM for thrombocytopenic samples; BC-6000 Plus-I (r = 0.966) was more relevant to IRM than any other method for samples with normal platelet counts. Correlation between Sysmex XN-9000-I (r = 0.960) and IRM was the highest among these methods for samples with thrombocytosis. For bias evaluation, the average bias of Sysmex XN-9000-F was -1.5 × 109/L (95% LA = -9.4 to +6.4) for samples with thrombocytopenia, compared with IRM. BC-6000 Plus-I had a small mean difference with IRM for samples with normal platelet counts or thrombocytosis. Moreover, all evaluated methods had acceptable sensitivity, specificity, and concordance rates as compared with IRM in the diagnosis of thrombocytopenia and thrombocytosis. Platelet counting by Sysmex XN-9000-F is more accurate than other methods for thrombocytopenic samples. BC-6000 Plus-I has superior association and consistency for normal platelet counts. As for thrombocytosis patients, Sysmex XN-9000-I has the highest correlation with IRM while Sysmex XN-9000-O has the highest diagnosis efficacy.

Identifiants

pubmed: 31125378
doi: 10.1371/journal.pone.0217298
pii: PONE-D-18-31826
pmc: PMC6534315
doi:

Types de publication

Comparative Study Evaluation Study Journal Article Research Support, Non-U.S. Gov't

Langues

eng

Sous-ensembles de citation

IM

Pagination

e0217298

Déclaration de conflit d'intérêts

The study was sponsored by Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Mindray provided BC-6000Plus hematology analyzer, reagents and other consumable materials, in addition to operational training for the instrument. The authors have declared that no competing interests exist. This does not alter our adherence to PLOS ONE policies on sharing data and materials.

Références

Vet Surg. 2015 Apr;44(3):304-13
pubmed: 25307726
Br J Haematol. 2018 Apr;181(2):267-270
pubmed: 28169426
Br J Haematol. 2005 Feb;128(4):520-5
pubmed: 15686462
Vascular. 2014 Dec;22(6):427-31
pubmed: 24522438
Transfus Apher Sci. 2001 Oct;25(2):93-106
pubmed: 11761280
Blood Coagul Fibrinolysis. 2015 Mar;26(2):235-7
pubmed: 25255243
Am J Clin Pathol. 1980 Aug;74(2):119-34
pubmed: 7405890
Int J Lab Hematol. 2007 Apr;29(2):77-91
pubmed: 17474881
Am J Clin Pathol. 2013 Oct;140(4):495-9
pubmed: 24045545
Thromb Haemost. 2016 Jun 2;115(6):1178-90
pubmed: 26842994
Am J Med Sci. 2016 Nov;352(5):502-506
pubmed: 27865298
Expert Rev Hematol. 2015 Oct;8(5):555-7
pubmed: 26174754
Cytometry B Clin Cytom. 2005 Sep;67(1):1-5
pubmed: 15918161
Sultan Qaboos Univ Med J. 2015 Nov;15(4):e463-8
pubmed: 26629371
Am J Clin Pathol. 2012 Jan;137(1):65-74
pubmed: 22180479
Expert Opin Drug Metab Toxicol. 2013 Dec;9(12):1667-75
pubmed: 24215532
Arch Pathol Lab Med. 2017 Jun;141(6):830-836
pubmed: 28402168
Haematologica. 2011 Oct;96(10):1462-9
pubmed: 21712540
Int J Lab Hematol. 2017 Jun;39(3):337-346
pubmed: 28263031
Hematology. 2001;5(6):421-7
pubmed: 27419344
BMC Vet Res. 2015 Nov 05;11:276
pubmed: 26542105
Am J Clin Pathol. 2010 Oct;134(4):634-47
pubmed: 20855645
Cytometry B Clin Cytom. 2010 Jul;78(4):279-85
pubmed: 20229505

Auteurs

Yi Sun (Y)

Department of Clinical Laboratory, Chinese People's Liberation Army General Hospital, Beijing, China.

Zuojian Hu (Z)

Department of Laboratory Medicine, First Affiliated Hospital of Guangxi Medical University, Nanning, Guangxi, China.

Zhili Huang (Z)

Department of Blood Transfusion, Second Affiliated Hospital of Guangxi Medical University, Nanning, Guangxi, China.

Huaping Chen (H)

Department of Laboratory Medicine, First Affiliated Hospital of Guangxi Medical University, Nanning, Guangxi, China.

Shanzi Qin (S)

Department of Laboratory Medicine, First Affiliated Hospital of Guangxi Medical University, Nanning, Guangxi, China.

Zhong Jianing (Z)

Department of Laboratory Medicine, First Affiliated Hospital of Guangxi Medical University, Nanning, Guangxi, China.

Siyuan Chen (S)

Department of Laboratory Medicine, First Affiliated Hospital of Guangxi Medical University, Nanning, Guangxi, China.

Xue Qin (X)

Department of Laboratory Medicine, First Affiliated Hospital of Guangxi Medical University, Nanning, Guangxi, China.

Yi Ye (Y)

Hematology Product Development Department, Shenzhen Mindray Bio-Medical Electronics Co., Ltd, Shenzhen, China.

Chengbin Wang (C)

Department of Clinical Laboratory, Chinese People's Liberation Army General Hospital, Beijing, China.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH