Safety and Efficacy Outcomes of Combined Leadless Pacemaker and Atrioventricular Nodal Ablation for Atrial Fibrillation Using a Single Femoral Puncture Approach.
Ablation
Atrial fibrillation
Atrioventricular node
Leadless pacemaker
Micra
Journal
Heart, lung & circulation
ISSN: 1444-2892
Titre abrégé: Heart Lung Circ
Pays: Australia
ID NLM: 100963739
Informations de publication
Date de publication:
May 2020
May 2020
Historique:
received:
16
01
2019
revised:
14
02
2019
accepted:
02
05
2019
pubmed:
19
6
2019
medline:
21
1
2021
entrez:
19
6
2019
Statut:
ppublish
Résumé
Atrioventricular nodal (AVN) ablation with permanent pacemaker (PPM) insertion is indicated for rate control in patients with atrial fibrillation (AF) who remain unresponsive to rate or rhythm control strategies. The leadless PPM (Micra Transcatheter Pacing System [TPS], Medtronic, Minneapolis, MN, USA) has the advantage of eliminating transvenous lead and pacemaker pocket-related complications. The aim of this case series was to determine the outcomes of patients who had undergone combined Micra TPS and AVN ablation, performed via a single femoral approach. A retrospective review was undertaken on patients who had undergone concurrent procedures, across two major hospitals in Perth, Western Australia. Procedural details were obtained from a cardiac devices database whilst patient demographics and clinical information were determined from medical records. Fourteen (14) patients underwent concurrent Micra TPS insertion and AVN ablation for symptomatic AF. The average age was 73±9.2 years, and 43% of them were males. There was no acute procedural/device related complication. Over a median follow-up duration of 9 months (36% completing 12-month follow-up), there was no incidence of device complications, in particular device dislodgement, malfunction or infection. One patient had a resuscitated ventricular fibrillation (VF) arrest event with new onset cardiomyopathy during follow-up and required Micra TPS removal. One patient died at 33 days post procedure from a non-cardiac cause. Device performance was excellent with stable sensing and pacing thresholds during the follow-up period. Our study has shown that combined leadless PPM (Micra TPS) implantation and AVN ablation using a single femoral approach is feasible, with good safety and efficacy profile in the short-medium term. Long-term data involving larger cohorts is needed to confirm the findings of this study and determine the clinical usefulness of this combined approach.
Sections du résumé
BACKGROUND
BACKGROUND
Atrioventricular nodal (AVN) ablation with permanent pacemaker (PPM) insertion is indicated for rate control in patients with atrial fibrillation (AF) who remain unresponsive to rate or rhythm control strategies. The leadless PPM (Micra Transcatheter Pacing System [TPS], Medtronic, Minneapolis, MN, USA) has the advantage of eliminating transvenous lead and pacemaker pocket-related complications. The aim of this case series was to determine the outcomes of patients who had undergone combined Micra TPS and AVN ablation, performed via a single femoral approach.
METHOD
METHODS
A retrospective review was undertaken on patients who had undergone concurrent procedures, across two major hospitals in Perth, Western Australia. Procedural details were obtained from a cardiac devices database whilst patient demographics and clinical information were determined from medical records.
RESULTS
RESULTS
Fourteen (14) patients underwent concurrent Micra TPS insertion and AVN ablation for symptomatic AF. The average age was 73±9.2 years, and 43% of them were males. There was no acute procedural/device related complication. Over a median follow-up duration of 9 months (36% completing 12-month follow-up), there was no incidence of device complications, in particular device dislodgement, malfunction or infection. One patient had a resuscitated ventricular fibrillation (VF) arrest event with new onset cardiomyopathy during follow-up and required Micra TPS removal. One patient died at 33 days post procedure from a non-cardiac cause. Device performance was excellent with stable sensing and pacing thresholds during the follow-up period.
CONCLUSION
CONCLUSIONS
Our study has shown that combined leadless PPM (Micra TPS) implantation and AVN ablation using a single femoral approach is feasible, with good safety and efficacy profile in the short-medium term. Long-term data involving larger cohorts is needed to confirm the findings of this study and determine the clinical usefulness of this combined approach.
Identifiants
pubmed: 31208898
pii: S1443-9506(19)30484-6
doi: 10.1016/j.hlc.2019.05.100
pii:
doi:
Types de publication
Journal Article
Multicenter Study
Langues
eng