High-throughput fluorescence polarization immunoassay by using a portable fluorescence polarization imaging analyzer.
Journal
Lab on a chip
ISSN: 1473-0189
Titre abrégé: Lab Chip
Pays: England
ID NLM: 101128948
Informations de publication
Date de publication:
07 08 2019
07 08 2019
Historique:
pubmed:
30
6
2019
medline:
30
6
2019
entrez:
29
6
2019
Statut:
ppublish
Résumé
High-throughput fluorescence polarization immunoassays (FPIAs) for mycotoxin were conducted using a portable FP analyzer with a microdevice. Simultaneous FPIA measurements for 8 different deoxynivalenol (DON) concentrations in 12 chambers (total of 96 samples) and high-throughput FPIA measurements for single DON concentrations in more than 500 chambers were conducted. The results indicated that simultaneous FPIAs for 96 independent samples and for 500 samples were possible by FP imaging. The FP analyzer has a size of 65 cm (W 35 cm × D 15 cm × H 15 cm) and costs less than $5000. The sample volume was 1 nL. Furthermore, it is expected that sample reaction and FP detection can be automatically conducted with the analyzer by changing the microdevice and the software. Its features such as low cost and portability will contribute to on-site measurement and point-of-care testing. Additionally, the high-throughput feature will contribute to the study of molecular interactions based on FP measurements.
Types de publication
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM