Intra- and inter-tester reliability of spasticity assessment in standing position in children and adolescents with cerebral palsy using a paediatric exoskeleton.
Lokomat
Psychometric properties
cerebral palsy
mobility
muscle tone
Journal
Disability and rehabilitation
ISSN: 1464-5165
Titre abrégé: Disabil Rehabil
Pays: England
ID NLM: 9207179
Informations de publication
Date de publication:
04 2021
04 2021
Historique:
pubmed:
2
8
2019
medline:
6
7
2021
entrez:
2
8
2019
Statut:
ppublish
Résumé
The L-STIFF tool of the Lokomat evaluates the hip and knee flexors and extensors spasticity in a standing position. It moves the lower limb at a controlled velocity, measuring joint resistance to passive movements. Since its reliability in children with cerebral palsy remains unknown, our goal was to evaluate the relative and absolute reliability of L-STIFF in children with cerebral palsy. Reliability was determined in 16 children with cerebral palsy by two experienced therapists. The changes in resistive torque in hip and knee in both flexion and extension were measured. Relative and absolute reliability were estimated using the intraclass correlation coefficient, standard error of measurement, and minimal detectable change. Reliability was assessed on three levels: (1) intra- and (2) inter-tester within session, and (3) intra-tester between sessions. Intraclass correlation coefficients were moderate to excellent for intra-tester reliability (all The assessment tool L-STIFF is a promising tool for quantifying lower limb spasticity in children with cerebral palsy in a standing position. However, the results should be interpreted carefully.Implications for RehabilitationL-STIFF is a promising tool for evaluating lower limb spasticity in standing position.A special care must be given to the installation and alignment of the participant into the Lokomat to minimize erroneous spasticity measurement.Relative standard error of measurement and minimal detectable change should be used to analyze changes spasticity.
Sections du résumé
BACKGROUND
The L-STIFF tool of the Lokomat evaluates the hip and knee flexors and extensors spasticity in a standing position. It moves the lower limb at a controlled velocity, measuring joint resistance to passive movements. Since its reliability in children with cerebral palsy remains unknown, our goal was to evaluate the relative and absolute reliability of L-STIFF in children with cerebral palsy.
METHODS
Reliability was determined in 16 children with cerebral palsy by two experienced therapists. The changes in resistive torque in hip and knee in both flexion and extension were measured. Relative and absolute reliability were estimated using the intraclass correlation coefficient, standard error of measurement, and minimal detectable change. Reliability was assessed on three levels: (1) intra- and (2) inter-tester within session, and (3) intra-tester between sessions.
RESULTS
Intraclass correlation coefficients were moderate to excellent for intra-tester reliability (all
CONCLUSIONS
The assessment tool L-STIFF is a promising tool for quantifying lower limb spasticity in children with cerebral palsy in a standing position. However, the results should be interpreted carefully.Implications for RehabilitationL-STIFF is a promising tool for evaluating lower limb spasticity in standing position.A special care must be given to the installation and alignment of the participant into the Lokomat to minimize erroneous spasticity measurement.Relative standard error of measurement and minimal detectable change should be used to analyze changes spasticity.