Experimental study on factors affecting the end effect in gas viscosity measurement using capillary-tube viscometer.
Journal
The Review of scientific instruments
ISSN: 1089-7623
Titre abrégé: Rev Sci Instrum
Pays: United States
ID NLM: 0405571
Informations de publication
Date de publication:
Jul 2019
Jul 2019
Historique:
entrez:
3
8
2019
pubmed:
3
8
2019
medline:
3
8
2019
Statut:
ppublish
Résumé
Because of its simple principle and high adaptability to severe operational conditions, the capillary-tube viscometer has been widely used for viscosity measurement. However, difficulties in accurately correcting the end effect induced measurement deviation will result in great uncertainty for measurement results. In order to solve this problem, in this work, we studied factors affecting the end effect by conducting the high pressure nitrogen viscosity measurement at low flow velocity with an improved capillary-tube viscometer. The experimental results indicated that the influence of the end effect became less significant with the decrease in flow velocity (v) and tube inner diameter (d) and varied inversely with the length of tube (L). We defined the ratio of measured viscosity to standard viscosity obtained from the NIST database as the viscosity deviation coefficient (C