Real-World Safety of an Implantable Continuous Glucose Sensor Over Multiple Cycles of Use: A Post-Market Registry Study.
Continuous glucose monitoring
Implantable
Safety
Sensor
Type 1 diabetes
Type 2 diabetes
Journal
Diabetes technology & therapeutics
ISSN: 1557-8593
Titre abrégé: Diabetes Technol Ther
Pays: United States
ID NLM: 100889084
Informations de publication
Date de publication:
01 2020
01 2020
Historique:
pubmed:
17
8
2019
medline:
6
11
2020
entrez:
17
8
2019
Statut:
ppublish
Résumé
Previously, the safety and accuracy of the Eversense continuous glucose monitoring (CGM) system were characterized in three pivotal trials among individuals with type 1 diabetes (T1D) and type 2 diabetes (T2D) with a single 90- or 180-day sensor insertion-removal cycle. The Post-Market Clinical Follow-up (PMCF) registry is a prospective study evaluating the long-term safety and performance of the Eversense CGM system over multiple sensor insertion-removal cycles among adults with T1D and T2D. All patients who had a sensor subcutaneously implanted across 534 participating centers in Europe and South Africa from June 2016 to August 2018 were enrolled. Adverse events (AEs) were recorded at each visit and patients were instructed to inform their clinic if they experienced any AEs between visits. AEs were adjudicated for relatedness to the device, procedure, or drug (dexamethasone acetate). The primary safety endpoint was the rate of related serious adverse events (SAEs) through four sensor insertion-removal cycles. The registry enrolled 3023 patients. As of last follow-up, 5417 sensors had been inserted with a total of 1260 patient-years (PYs) of follow-up: 969 patients had used the system for at least 6 months and 173 patients had used the system for at least 1 year. No related SAEs were reported. The most frequently reported related AEs were sensor location site infection (0.96%; 2.46 events per 100 PYs), inability to remove the sensor upon first attempt (0.76%; 1.90 events per 100 PYs), and adhesive patch location site irritation (0.66%; 1.59 events per 100 PYs). One nonserious allergic reaction to lidocaine was reported, which resolved with administration of an antihistamine. The full intended sensor life was achieved by 91% of 90-day sensors and 75% of 180-day sensors. The PMCF registry provides real-world evidence that the Eversense CGM system is safe over multiple cycles of use.
Identifiants
pubmed: 31418587
doi: 10.1089/dia.2019.0159
pmc: PMC6945795
doi:
Substances chimiques
Blood Glucose
0
Types de publication
Journal Article
Multicenter Study
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM
Pagination
48-52Références
Diabetes Care. 1995 Mar;18(3):361-76
pubmed: 7555480
Diabetes Care. 2014 Oct;37(10):2702-9
pubmed: 25011947
Clin Diabetes. 2018 Jan;36(1):50-58
pubmed: 29382979
JAMA. 2017 Jan 24;317(4):371-378
pubmed: 28118453
Diabetes Technol Ther. 2019 May;21(5):231-237
pubmed: 30925083
Diabetes Technol Ther. 2019 Feb;21(2):66-72
pubmed: 30657336
JAMA. 2017 Jan 24;317(4):379-387
pubmed: 28118454
Ann Intern Med. 2017 Sep 19;167(6):365-374
pubmed: 28828487
Acta Diabetol. 2014 Oct;51(5):845-51
pubmed: 25037251
Diabetes Technol Ther. 2018 Mar;20(3):197-206
pubmed: 29381090
Contact Dermatitis. 2017 Dec;77(6):367-373
pubmed: 28804907
Diabetes Care. 2017 Jan;40(1):63-68
pubmed: 27815290