Fan-Beam Based Virtual Fluoroscopy for Navigated Catheterization in Interventional Radiology.


Journal

Studies in health technology and informatics
ISSN: 1879-8365
Titre abrégé: Stud Health Technol Inform
Pays: Netherlands
ID NLM: 9214582

Informations de publication

Date de publication:
21 Aug 2019
Historique:
entrez: 24 8 2019
pubmed: 24 8 2019
medline: 5 9 2019
Statut: ppublish

Résumé

Personalized medicine implies reducing invasiveness of therapeutic procedures. Although interventional radiology proved a very interesting alternative to surgical procedures, it still raises concerns due to the irradiation dose received by the medical team (and by the patient). We propose a novel concept allowing to reduce very significantly the irradiation dose during the phases where tools inserted in the patient have to be tracked with respect to previously acquired images. This implies inserting a miniaturized X-ray detector in the tip of the tools, and reducing the dose by a "rotating collimator". We demonstrate that real-time processing of the signals allows accurate localization of the tip of the tools, with a dose reduction of at least ten times.

Identifiants

pubmed: 31437888
pii: SHTI190186
doi: 10.3233/SHTI190186
doi:

Types de publication

Journal Article

Langues

eng

Pagination

74-78

Auteurs

Pierrick Guiral (P)

Univ. Grenoble Alpes, CNRS, Grenoble INP, TIMC-IMAG, F-38000 Grenoble, France.

Patrick Pittet (P)

Université de Lyon, CNRS, Université Lyon 1, INL, F-69100 Villeurbanne, France.

Yannick Grondin (Y)

SurgiQual Institute, F-38240 Meylan, France.

Patrice Jalade (P)

HCL, Centre Hospitalier Lyon Sud, Service de Radiophysique et Radiovigilance, F-69310 Pierre-Bénite, France.

Jean-Marc Galvan (JM)

Université de Lyon, CNRS, Université Lyon 1, INL, F-69100 Villeurbanne, France.

Guo-Neng Lu (GN)

Université de Lyon, CNRS, Université Lyon 1, INL, F-69100 Villeurbanne, France.

Laurent Desbat (L)

Univ. Grenoble Alpes, CNRS, Grenoble INP, TIMC-IMAG, F-38000 Grenoble, France.

Philippe Cinquin (P)

Univ. Grenoble Alpes, CNRS, Grenoble INP, TIMC-IMAG, F-38000 Grenoble, France.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH