Monolithic chromatography on conjoint liquid chromatography columns for speciation of platinum-based chemotherapeutics in serum of cancer patients.
Anion-exchange and affinity monolithic disks
Inductively coupled plasma mass spectrometry
Low-pressure and high-pressure conjoint liquid chromatography
Pt-based chemotherapeutics
Serum of cancer patients
Journal
Journal of trace elements in medicine and biology : organ of the Society for Minerals and Trace Elements (GMS)
ISSN: 1878-3252
Titre abrégé: J Trace Elem Med Biol
Pays: Germany
ID NLM: 9508274
Informations de publication
Date de publication:
Jan 2020
Jan 2020
Historique:
received:
13
04
2019
revised:
14
09
2019
accepted:
17
09
2019
pubmed:
27
9
2019
medline:
21
4
2020
entrez:
27
9
2019
Statut:
ppublish
Résumé
Monolithic chromatography using convective interaction media (CIM) disks or columns can be used in the separation step of speciation analysis. When different monolithic disks are placed in one housing, forming conjoint liquid chromatography (CLC) monolithic column, two-dimensional separation is achieved in a single chromatographic run. Here, we assembled low-pressure (maximum 50 bar) CLC monolithic column, which consists of two 0.34 mL shallow CIM monolithic disks and high-pressure CLC column (maximum 150 bar) from 0.1 mL analytical high performance short bed CIMac monolithic disks. Both the CLC columns constructed from affinity Protein G and weak anion exchange diethylamine (DEAE) disks, were applied for the speciation of cisplatin, oxaliplatin and carboplatin in spiked standard serum proteins, spiked human serum and serum of cancer patients. The analytical performances of the CLC columns used were evaluated by comparing their robustness, selectivity, repeatability and reproducibility. The separated serum proteins were detected on-line by ultraviolet (UV) and eluted Pt species by inductively coupled plasma mass spectrometry (ICP-MS). For accurate quantification of the separated Pt species (unbound Pt-based chemotherapeutic from species associated to transferrin (Tf), human serum albumin (HSA) and Immunoglobulin G (IgG)), post column isotope dilution (ID)-ICP-MS was used. The data from analyses showed that both tested CLC monolithic columns gave statistically comparable results, with the low-pressure CLC column exhibiting better resolving power and robustness. It also enables more effective cleaning of monolithic disks and to analyse larger series of serum samples than the high-pressure CLC column. Analyses of serum samples of cancer patients treated with cisplatin or carboplatin showed that Pt-chemotherapeutics were bound preferentially to HSA (around 80%). The portion of unbound Pt in general did not exceed 2%, up to 5% of Pt was associated with Tf and approximately 20% with IgG. Column recoveries, calculated as a ratio between the sum of concentrations of Pt species eluted and concentration of total Pt in serum samples, were close to 100%. Low-pressure CLC column exhibited greater potential than high-pressure CLC column, and can be thus recommended for its intended use in speciation analysis of metal-based biomolecules.
Sections du résumé
BACKGROUND
BACKGROUND
Monolithic chromatography using convective interaction media (CIM) disks or columns can be used in the separation step of speciation analysis. When different monolithic disks are placed in one housing, forming conjoint liquid chromatography (CLC) monolithic column, two-dimensional separation is achieved in a single chromatographic run.
METHODS
METHODS
Here, we assembled low-pressure (maximum 50 bar) CLC monolithic column, which consists of two 0.34 mL shallow CIM monolithic disks and high-pressure CLC column (maximum 150 bar) from 0.1 mL analytical high performance short bed CIMac monolithic disks. Both the CLC columns constructed from affinity Protein G and weak anion exchange diethylamine (DEAE) disks, were applied for the speciation of cisplatin, oxaliplatin and carboplatin in spiked standard serum proteins, spiked human serum and serum of cancer patients. The analytical performances of the CLC columns used were evaluated by comparing their robustness, selectivity, repeatability and reproducibility. The separated serum proteins were detected on-line by ultraviolet (UV) and eluted Pt species by inductively coupled plasma mass spectrometry (ICP-MS). For accurate quantification of the separated Pt species (unbound Pt-based chemotherapeutic from species associated to transferrin (Tf), human serum albumin (HSA) and Immunoglobulin G (IgG)), post column isotope dilution (ID)-ICP-MS was used.
RESULTS
RESULTS
The data from analyses showed that both tested CLC monolithic columns gave statistically comparable results, with the low-pressure CLC column exhibiting better resolving power and robustness. It also enables more effective cleaning of monolithic disks and to analyse larger series of serum samples than the high-pressure CLC column. Analyses of serum samples of cancer patients treated with cisplatin or carboplatin showed that Pt-chemotherapeutics were bound preferentially to HSA (around 80%). The portion of unbound Pt in general did not exceed 2%, up to 5% of Pt was associated with Tf and approximately 20% with IgG. Column recoveries, calculated as a ratio between the sum of concentrations of Pt species eluted and concentration of total Pt in serum samples, were close to 100%.
CONCLUSIONS
CONCLUSIONS
Low-pressure CLC column exhibited greater potential than high-pressure CLC column, and can be thus recommended for its intended use in speciation analysis of metal-based biomolecules.
Identifiants
pubmed: 31557573
pii: S0946-672X(19)30246-9
doi: 10.1016/j.jtemb.2019.09.011
pii:
doi:
Substances chimiques
Immunoglobulin G
0
Serum Albumin
0
Platinum
49DFR088MY
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
28-39Informations de copyright
Copyright © 2019 Elsevier GmbH. All rights reserved.