Comparison of estimation of cardiac output using an uncalibrated pulse contour method and echocardiography during veno-venous extracorporeal membrane oxygenation.
cardiac output
extracorporeal membrane oxygenation
pulse contour method
respiratory failure
Journal
Perfusion
ISSN: 1477-111X
Titre abrégé: Perfusion
Pays: England
ID NLM: 8700166
Informations de publication
Date de publication:
07 2020
07 2020
Historique:
pubmed:
12
11
2019
medline:
29
5
2021
entrez:
12
11
2019
Statut:
ppublish
Résumé
During veno-venous extracorporeal membrane oxygenation, cardiac output monitoring is essential to assess tissue oxygen delivery. Adequate arterial oxygenation depends on the ratio between the extracorporeal pump blood flow and the cardiac output. The aim of this study was to compare estimates of cardiac output and blood flow/cardiac output ratios made using an uncalibrated pulse contour method with those made using echocardiography in patients treated with veno-venous extracorporeal membrane oxygenation. Cardiac output was estimated simultaneously using a pulse contour method (MostCareUp; Vygon, Encouen, France) and echocardiography in 17 hemodynamically stable patients treated with veno-venous extracorporeal membrane oxygenation. Comparisons were made using Bland-Altman and linear regression analysis. There were significant correlations between cardiac output estimated using pulse contour method and echocardiography and between blood flow/cardiac output estimated using pulse contour method and blood flow/cardiac output estimated using echocardiography (r = 0.84, p < 0.001 and r = 0.87, p < 0.001, respectively). Bland-Altman analysis showed a good agreement (bias -0.20 ± 0.50 L/min) and a low percentage of error (25%) for the cardiac output values estimated by the two methods. The bias between the blood flow/cardiac output ratios obtained with the two methods was 5.19% ± 12.3% (percentage of error = 28.1%). The pulse contour method is a valuable alternative to echocardiography for the assessment of cardiac output and the blood flow/cardiac output ratio in patients treated with veno-venous extracorporeal membrane oxygenation.
Identifiants
pubmed: 31709902
doi: 10.1177/0267659119883204
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM