Patients with pacemakers or defibrillators do not need to worry about e-Cars: An observational study.
Electric cars
cardiac implantable electronic devices (CIED)
defibrillator
electromagnetic interference (EMI)
pacemaker
Journal
Technology and health care : official journal of the European Society for Engineering and Medicine
ISSN: 1878-7401
Titre abrégé: Technol Health Care
Pays: Netherlands
ID NLM: 9314590
Informations de publication
Date de publication:
2020
2020
Historique:
pubmed:
21
11
2019
medline:
25
11
2020
entrez:
21
11
2019
Statut:
ppublish
Résumé
Electric cars are increasingly used for public and private transportation and represent possible sources of electromagnetic interference (EMI). Potential implications for patients with cardiac implantable electronic devices (CIED) range from unnecessary driving restrictions to life-threatening device malfunction. This prospective, cross-sectional study was designed to assess the EMI risk of electric cars on CIED function. One hundred and eight consecutive patients with CIEDs presenting for routine follow-up between May 2014 and January 2015 were enrolled in the study. The participants were exposed to electromagnetic fields generated by the four most common electric cars (Nissan Leaf, Tesla Model S, BMW i3, VW eUp) while roller-bench test-driving at Institute of Automotive Technology, Department of Mechanical Engineering, Technical University, Munich. The primary endpoint was any abnormalities in CIED function (e.g. oversensing with pacing-inhibition, inappropriate therapy or mode-switching) while driving or charging electric cars as assessed by electrocardiographic recordings and device interrogation. No change in device function or programming was seen in this cohort which is representative of contemporary CIED devices. The largest electromagnetic field detected was along the charging cable during high current charging (116.5 μT). The field strength in the cabin was lower (2.1-3.6 μT). Electric cars produce electromagnetic fields; however, they did not affect CIED function or programming in our cohort. Driving and charging of electric cars is likely safe for patients with CIEDs.
Sections du résumé
BACKGROUND
BACKGROUND
Electric cars are increasingly used for public and private transportation and represent possible sources of electromagnetic interference (EMI). Potential implications for patients with cardiac implantable electronic devices (CIED) range from unnecessary driving restrictions to life-threatening device malfunction. This prospective, cross-sectional study was designed to assess the EMI risk of electric cars on CIED function.
METHODS
METHODS
One hundred and eight consecutive patients with CIEDs presenting for routine follow-up between May 2014 and January 2015 were enrolled in the study. The participants were exposed to electromagnetic fields generated by the four most common electric cars (Nissan Leaf, Tesla Model S, BMW i3, VW eUp) while roller-bench test-driving at Institute of Automotive Technology, Department of Mechanical Engineering, Technical University, Munich. The primary endpoint was any abnormalities in CIED function (e.g. oversensing with pacing-inhibition, inappropriate therapy or mode-switching) while driving or charging electric cars as assessed by electrocardiographic recordings and device interrogation.
RESULTS
RESULTS
No change in device function or programming was seen in this cohort which is representative of contemporary CIED devices. The largest electromagnetic field detected was along the charging cable during high current charging (116.5 μT). The field strength in the cabin was lower (2.1-3.6 μT).
CONCLUSIONS
CONCLUSIONS
Electric cars produce electromagnetic fields; however, they did not affect CIED function or programming in our cohort. Driving and charging of electric cars is likely safe for patients with CIEDs.