Optimal Site for Proximal Optimization Technique in Complex Coronary Bifurcation Stenting: A Computational Fluid Dynamics Study.


Journal

Cardiovascular revascularization medicine : including molecular interventions
ISSN: 1878-0938
Titre abrégé: Cardiovasc Revasc Med
Pays: United States
ID NLM: 101238551

Informations de publication

Date de publication:
07 2020
Historique:
received: 01 11 2019
revised: 22 11 2019
accepted: 09 12 2019
pubmed: 24 12 2019
medline: 26 1 2021
entrez: 24 12 2019
Statut: ppublish

Résumé

The optimal position of the balloon distal radio-opaque marker during the post optimization technique (POT) remains debated. We analyzed three potential different balloon positions for the final POT in two different two-stenting techniques, to compare the hemodynamic effects in terms of wall shear stress (WSS) in patients with complex left main (LM) coronary bifurcation. We reconstructed the patient-specific coronary bifurcation anatomy using the coronary computed tomography angiography (CCTA) data of 8 consecutive patients (6 males, mean age 68.2± 18.6 years) affected by complex LM bifurcation disease. Subsequently a virtual bench test was performed in each patient using two different double stenting techniques represented by the DK and Nano crush using the reconstruction of Orsiro stents (Biotronik IC, Bulack, Switzerland). A significant reduction in the mean WSS values in all the lesion's sites was observed when the final POT was performed 1 mm distally the carina cut plane in both techniques. Moreover, a significant improvement in the mean WSS values of the entire SB (e.g. LCX) was obtained performing the POT 1 mm distally to the carina cut plane. The proximal POT resulted in larger area of lower WSS values at the carina using both the Nano crush and the DK crush techniques. In patients with complex LM bifurcation disease the use of a final POT performed 1 mm distally to the carina cut plane might results in more favorable WSS patterns (i.e. higher WSS values) along all stented segments and, especially, along the entire LCX lesions.

Sections du résumé

BACKGROUND/PURPOSE
The optimal position of the balloon distal radio-opaque marker during the post optimization technique (POT) remains debated. We analyzed three potential different balloon positions for the final POT in two different two-stenting techniques, to compare the hemodynamic effects in terms of wall shear stress (WSS) in patients with complex left main (LM) coronary bifurcation.
METHODS/MATERIALS
We reconstructed the patient-specific coronary bifurcation anatomy using the coronary computed tomography angiography (CCTA) data of 8 consecutive patients (6 males, mean age 68.2± 18.6 years) affected by complex LM bifurcation disease. Subsequently a virtual bench test was performed in each patient using two different double stenting techniques represented by the DK and Nano crush using the reconstruction of Orsiro stents (Biotronik IC, Bulack, Switzerland).
RESULTS
A significant reduction in the mean WSS values in all the lesion's sites was observed when the final POT was performed 1 mm distally the carina cut plane in both techniques. Moreover, a significant improvement in the mean WSS values of the entire SB (e.g. LCX) was obtained performing the POT 1 mm distally to the carina cut plane. The proximal POT resulted in larger area of lower WSS values at the carina using both the Nano crush and the DK crush techniques.
CONCLUSIONS
In patients with complex LM bifurcation disease the use of a final POT performed 1 mm distally to the carina cut plane might results in more favorable WSS patterns (i.e. higher WSS values) along all stented segments and, especially, along the entire LCX lesions.

Identifiants

pubmed: 31866275
pii: S1553-8389(19)30811-5
doi: 10.1016/j.carrev.2019.12.015
pii:
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

826-832

Commentaires et corrections

Type : CommentIn

Informations de copyright

Copyright © 2019 Elsevier Inc. All rights reserved.

Déclaration de conflit d'intérêts

Declaration of competing interest None of the authors have conflict of interest to declare.

Auteurs

Marco Zuin (M)

Section of Internal and Cardiopulmonary Medicine, University of Ferrara, Faculty of Medicine Ferrara, Italy; Department of Cardiovascular Diagnosis and Endoluminal Interventions, Santa Maria della Misericordia Hospital, Rovigo, Italy.

Gianluca Rigatelli (G)

Department of Cardiovascular Diagnosis and Endoluminal Interventions, Santa Maria della Misericordia Hospital, Rovigo, Italy. Electronic address: jackyheart@libero.it.

Claudio Chiastra (C)

PoliToBIOMed Lab, Department of Mechanical and Aerospace Engineering, Politecnico di Torino, Turin, Italy.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH