To describe a method for averaging ocular coherence tomography angiography (OCTA) images using a commercially available photo editing software: Adobe Photoshop CC 2017. This single-center study assessed the feasibility of using Adobe Photoshop as an OCTA image averaging tool. Three 3.0 mm × 3.0 mm OCTA images from each eye were obtained using commercially available OCTA devices. Captured OCTA images were exported in high-resolution TIFF format, imported as an image stack, aligned using an automated function, and averaged by creating a Smart Object using Photoshop CC 2017 software. In conjunction with qualitative assessment, the main outcome of the study was image grader preferences with respect to clarity of the foveal avascular zone (FAZ), blood vessel delineation, and the ability to identify abnormal vasculature. After removing OCTA scans with significant image distortion, 25 sets of images were included in the analysis. Adobe Photoshop CC 2017 successfully aligned and averaged all images of the superficial and deep retinal plexuses that contained a minimum 40% overlap. Three independent retinal specialists found the averaged images to be slightly or definitely preferable to the original 87%, 89%, and 69% of the time with respect to clarity of the FAZ, clarity of blood vessel delineation, and the ability to identify abnormal vasculature, respectively. Adobe Photoshop CC 2017 is an excellent tool for image averaging, producing high-quality resulting OCTA images. As an easily accessible software, Photoshop has the potential for use in a diversity of pathological conditions. [Ophthalmic Surg Lasers Imaging Retina. 2019;50:802-807.].
Sections du résumé
BACKGROUND AND OBJECTIVE
To describe a method for averaging ocular coherence tomography angiography (OCTA) images using a commercially available photo editing software: Adobe Photoshop CC 2017.
PATIENTS AND METHODS
This single-center study assessed the feasibility of using Adobe Photoshop as an OCTA image averaging tool. Three 3.0 mm × 3.0 mm OCTA images from each eye were obtained using commercially available OCTA devices. Captured OCTA images were exported in high-resolution TIFF format, imported as an image stack, aligned using an automated function, and averaged by creating a Smart Object using Photoshop CC 2017 software. In conjunction with qualitative assessment, the main outcome of the study was image grader preferences with respect to clarity of the foveal avascular zone (FAZ), blood vessel delineation, and the ability to identify abnormal vasculature.
RESULTS
After removing OCTA scans with significant image distortion, 25 sets of images were included in the analysis. Adobe Photoshop CC 2017 successfully aligned and averaged all images of the superficial and deep retinal plexuses that contained a minimum 40% overlap. Three independent retinal specialists found the averaged images to be slightly or definitely preferable to the original 87%, 89%, and 69% of the time with respect to clarity of the FAZ, clarity of blood vessel delineation, and the ability to identify abnormal vasculature, respectively.
CONCLUSIONS
Adobe Photoshop CC 2017 is an excellent tool for image averaging, producing high-quality resulting OCTA images. As an easily accessible software, Photoshop has the potential for use in a diversity of pathological conditions. [Ophthalmic Surg Lasers Imaging Retina. 2019;50:802-807.].