A miniaturized accelerometer attached to the heart has been used for monitoring functional parameters such as early systolic velocity and displacement. Currently, processing of the accelerometer signal for derival of these functional parameters depends on determining start systole by detecting the ECG R-peaks. This study proposes an alternative method using only the accelerometer signal to detect start systole, making additional ECG recordings for this purpose redundant. A signal processing method for automatic detection of start systole by accelerometer alone was developed and compared with detected R-peaks in 15 pigs during 5 different interventions showing a difference of 30 ± 17 ms. Furthermore, the derived early systolic velocity and displacement using only accelerometer measurements correlated well (r