Surface Dosimetry of Patients Undergoing Total Body Irradiation: A Retrospective Analysis for Quality Assurance.
bone marrow transplantation
clinical dosimetry data
quality assurance
total body irradiation
Journal
Cureus
ISSN: 2168-8184
Titre abrégé: Cureus
Pays: United States
ID NLM: 101596737
Informations de publication
Date de publication:
06 Feb 2020
06 Feb 2020
Historique:
entrez:
18
2
2020
pubmed:
18
2
2020
medline:
18
2
2020
Statut:
epublish
Résumé
Total body irradiation (TBI) is used prior to bone marrow transplantation as part of the conditioning regimen in selected patients. A linear accelerator-based technique was used at our treatment centre between June, 2004 and August, 2015. Patients were treated supine with extended source-to-surface distance (SSD) lateral fields, and prescription dose was 12 Gy delivered in six fractions, two fractions per day. Dose was prescribed to midplane at the level of the umbilicus and monitor units were calculated manually based on measured beam data. Dose variation within 10% of the prescribed midplane dose is considered acceptable for TBI treatment. This was achieved in our clinic by using compensators to account for missing tissue in the head and neck and lower leg regions. Lung attenuators were routinely used to correct for internal inhomogeneity, which resulted from low density lung tissue. The purpose of this study was to determine whether dose variation was within acceptable limits for these patients as part of a quality assurance process. Following chart review, 129 patients who received six-fraction TBI from 2004 to 2015 were included in this study. Patients receiving single fraction treatment were excluded. Metal oxide semiconductor field effect transistors (MOSFET) dosimetry was used to measure surface dose at four or five locations during patients' first fraction of TBI. Dosimetry was repeated during the second fraction for any site with variation greater than 10%. Statistical analysis was carried out on patient data, diagnosis and dosimetry measurements. Of the 129 patients who met the inclusion criteria, 50 were diagnosed with acute myelogenous leukemia, 30 with acute lymphoblastic leukemia and 11 with chronic myelogenous leukemia. The rest of the patients were diagnosed with lymphoma or myelodysplastic syndromes. The mean percent variation in dosimetry measurements taken at the specific locations ranged between 3.5% and 8.3%. The highest variation was found in measurements performed on the cheek. A high percentage of all dosimetry readings (85.5%) was within the acceptable range of +10% from the expected value. The highest number of individual readings taken at a specific location that fell outside this range were found at the cheek. We conclude that the linear accelerator delivered TBI at our centre meets the acceptable limits of dose variation over an 11-year period.
Identifiants
pubmed: 32064218
doi: 10.7759/cureus.6900
pmc: PMC7003717
doi:
Types de publication
Journal Article
Langues
eng
Pagination
e6900Informations de copyright
Copyright © 2020, Sengupta et al.
Déclaration de conflit d'intérêts
The authors have declared that no competing interests exist.
Références
Med Phys. 1994 Apr;21(4):567-72
pubmed: 8058024
Phys Med Biol. 1996 Nov;41(11):2327-43
pubmed: 8938029
Radiat Prot Dosimetry. 2002;101(1-4):393-8
pubmed: 12382775
Phys Med Biol. 2004 Sep 7;49(17):4031-48
pubmed: 15470921
Int J Radiat Oncol Biol Phys. 2007 Nov 15;69(4):1297-304
pubmed: 17881143
J Appl Clin Med Phys. 2014 Jul 08;15(4):4939
pubmed: 25207423