Acute in vivo testing of a polymer cuff electrode with integrated microfluidic channels for stimulation, recording, and drug delivery on rat sciatic nerve.
Cuff electrode
Drug delivery
Electrophysiology
Peripheral nerve
Journal
Journal of neuroscience methods
ISSN: 1872-678X
Titre abrégé: J Neurosci Methods
Pays: Netherlands
ID NLM: 7905558
Informations de publication
Date de publication:
15 04 2020
15 04 2020
Historique:
received:
03
12
2019
revised:
11
02
2020
accepted:
11
02
2020
pubmed:
19
2
2020
medline:
22
6
2021
entrez:
19
2
2020
Statut:
ppublish
Résumé
Extraneural cuffs are among the least invasive peripheral nerve interfaces as they remain outside the nerve. However, compared with more invasive interfaces, these electrodes may suffer from lower selectivity and sensitivity since the targeted nerve fibers are more distanced from the electrodes. A lyse-and-attract cuff electrode (LACE) was enabled by microfabrication and developed to improve selectivity and sensitivity while maintaining a cuff format. Its engineering design was described in previous work. LACE is a hybrid cuff that integrates both microelectrodes and microfluidic channels. The ultimate goal is to increase fascicular selectivity and sensitivity by focal delivery via the microchannels of (1) lysing agent to remove connective tissue separating electrodes from nerve fibers, and (2) neurotrophic factors to promote axonal sprouting of the exposed nerve fibers into microfluidic channels where electrodes are embedded. Here, we focus on demonstrating in vivo function of microfluidics and microelectrodes in an acute preparation in which we evaluate the ability to focally remove connective tissue and record and stimulate with microchannel-embedded microelectrodes neural activity in rat sciatic nerves. While extraneural interfaces prioritize nerve health and intraneural interfaces prioritize functionality, LACE represents a new extraneural approach which could potentially excel at both aims. Surgical implantation demonstrate preservation of LACE function following careful and minimal handling. In vivo electrical evaluation demonstrates the ability of microelectrodes placed within microfluidic channels to successfully stimulate and record compound action potentials from rat sciatic nerve. Furthermore, collagen-rich epineurium was focally removed following infusion of collagenase via microchannels and confirmed via microscopy. The feasibility of using a cuff having integrated microelectrodes and microfluidics to stimulate, record, and deliver drug to focally lyse away the epineurium layer was demonstrated in acute experiments on rat sciatic nerve.
Sections du résumé
BACKGROUND
Extraneural cuffs are among the least invasive peripheral nerve interfaces as they remain outside the nerve. However, compared with more invasive interfaces, these electrodes may suffer from lower selectivity and sensitivity since the targeted nerve fibers are more distanced from the electrodes.
NEW METHOD
A lyse-and-attract cuff electrode (LACE) was enabled by microfabrication and developed to improve selectivity and sensitivity while maintaining a cuff format. Its engineering design was described in previous work. LACE is a hybrid cuff that integrates both microelectrodes and microfluidic channels. The ultimate goal is to increase fascicular selectivity and sensitivity by focal delivery via the microchannels of (1) lysing agent to remove connective tissue separating electrodes from nerve fibers, and (2) neurotrophic factors to promote axonal sprouting of the exposed nerve fibers into microfluidic channels where electrodes are embedded. Here, we focus on demonstrating in vivo function of microfluidics and microelectrodes in an acute preparation in which we evaluate the ability to focally remove connective tissue and record and stimulate with microchannel-embedded microelectrodes neural activity in rat sciatic nerves.
COMPARISON WITH EXISTING METHODS
While extraneural interfaces prioritize nerve health and intraneural interfaces prioritize functionality, LACE represents a new extraneural approach which could potentially excel at both aims.
RESULTS
Surgical implantation demonstrate preservation of LACE function following careful and minimal handling. In vivo electrical evaluation demonstrates the ability of microelectrodes placed within microfluidic channels to successfully stimulate and record compound action potentials from rat sciatic nerve. Furthermore, collagen-rich epineurium was focally removed following infusion of collagenase via microchannels and confirmed via microscopy.
CONCLUSION
The feasibility of using a cuff having integrated microelectrodes and microfluidics to stimulate, record, and deliver drug to focally lyse away the epineurium layer was demonstrated in acute experiments on rat sciatic nerve.
Identifiants
pubmed: 32068010
pii: S0165-0270(20)30056-X
doi: 10.1016/j.jneumeth.2020.108634
pii:
doi:
Substances chimiques
Pharmaceutical Preparations
0
Polymers
0
Types de publication
Journal Article
Research Support, Non-U.S. Gov't
Research Support, U.S. Gov't, Non-P.H.S.
Langues
eng
Sous-ensembles de citation
IM
Pagination
108634Commentaires et corrections
Type : ErratumIn
Informations de copyright
Published by Elsevier B.V.