Impact of hydrostatic pressure on fractional flow reserve: in vivo experimental study of anatomical height difference of coronary arteries.
Cardiovascular intervention
Fractional flow reserve
Hydrostatic pressure
Ischemic heart disease
Journal
Journal of cardiology
ISSN: 1876-4738
Titre abrégé: J Cardiol
Pays: Netherlands
ID NLM: 8804703
Informations de publication
Date de publication:
07 2020
07 2020
Historique:
received:
11
09
2019
revised:
02
12
2019
accepted:
20
01
2020
pubmed:
25
2
2020
medline:
18
5
2021
entrez:
25
2
2020
Statut:
ppublish
Résumé
Although pressure equalization of the sensor-tipped guidewire and systemic pressure is mandatory in measuring fractional flow reserve (FFR), pressure in the distal artery (Pd) with wire advancement can be influenced by hydrostatic pressure related to the height difference between the catheter tip and the distal pressure sensor. We therefore analyzed the impact of hydrostatic pressure on FFR in vivo by modification of the height difference. To reveal the anatomical height difference in human coronary arteries, measurement was performed during computed tomography angiography (CTA) of five consecutive patients. Utilizing the healthy coronary arteries of female swine, height difference diversity was reproduced by body rotation and vertical inclination. FFR measurements were performed during maximum hyperemia with adenosine. The height difference was calculated fluoroscopically with a contrast medium-filled balloon for reference. In human coronary CTA, height averages from the ostium in the left anterior descending artery (34.6 mm) were significantly higher than in the left circumflex (-15.5 mm, p = 0.008) and right coronary arteries (-2.3 mm, p = 0.008). In our swine model, reproduced height variation ranged from -7.2 cm to +6.5 cm. Mean FFR was significantly lower in positive sensor height and higher in negative sensor height compared to the reference height. Linear regression analyses revealed significant correlations between height difference and FFR, observed among all coronary arteries, as well as between the height difference and Pd-aortic pressure mismatch. Subtracting 0.622 mmHg/cm height difference from Pd could correct the expected hydrostatic pressure influence. Hydrostatic pressure variation resulting from sensor height influenced FFR values might affect interpretation during FFR assessment.
Sections du résumé
BACKGROUND
Although pressure equalization of the sensor-tipped guidewire and systemic pressure is mandatory in measuring fractional flow reserve (FFR), pressure in the distal artery (Pd) with wire advancement can be influenced by hydrostatic pressure related to the height difference between the catheter tip and the distal pressure sensor. We therefore analyzed the impact of hydrostatic pressure on FFR in vivo by modification of the height difference.
METHODS
To reveal the anatomical height difference in human coronary arteries, measurement was performed during computed tomography angiography (CTA) of five consecutive patients. Utilizing the healthy coronary arteries of female swine, height difference diversity was reproduced by body rotation and vertical inclination. FFR measurements were performed during maximum hyperemia with adenosine. The height difference was calculated fluoroscopically with a contrast medium-filled balloon for reference.
RESULTS
In human coronary CTA, height averages from the ostium in the left anterior descending artery (34.6 mm) were significantly higher than in the left circumflex (-15.5 mm, p = 0.008) and right coronary arteries (-2.3 mm, p = 0.008). In our swine model, reproduced height variation ranged from -7.2 cm to +6.5 cm. Mean FFR was significantly lower in positive sensor height and higher in negative sensor height compared to the reference height. Linear regression analyses revealed significant correlations between height difference and FFR, observed among all coronary arteries, as well as between the height difference and Pd-aortic pressure mismatch. Subtracting 0.622 mmHg/cm height difference from Pd could correct the expected hydrostatic pressure influence.
CONCLUSION
Hydrostatic pressure variation resulting from sensor height influenced FFR values might affect interpretation during FFR assessment.
Identifiants
pubmed: 32089479
pii: S0914-5087(20)30028-9
doi: 10.1016/j.jjcc.2020.01.009
pii:
doi:
Types de publication
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM
Pagination
73-79Commentaires et corrections
Type : CommentIn
Type : CommentIn
Informations de copyright
Copyright © 2020. Published by Elsevier Ltd.