Evaluation of Efficacy of a New Cryoprobe for Transbronchial Cryobiopsy: A Randomized, Controlled in vivo Animal Study.


Journal

Respiration; international review of thoracic diseases
ISSN: 1423-0356
Titre abrégé: Respiration
Pays: Switzerland
ID NLM: 0137356

Informations de publication

Date de publication:
Historique:
received: 03 09 2019
accepted: 18 01 2020
pubmed: 27 2 2020
medline: 10 4 2021
entrez: 27 2 2020
Statut: ppublish

Résumé

Forceps biopsy (FB) is still the most popular procedure for the bronchoscopic sampling of lung tissue. However, it has limitations like inadequate sample size and crush artifacts. Cryobiopsy (CB) has been introduced to obtain bronchoscopic biopsies with improved diagnostic yield compared to FB. Limitations of CB are the need to retract the cryoprobe en bloc with bronchoscope because samples are larger than the working channel and the variations of the freezing power of the reusable CB (rCB). Therefore, 3 new disposable cryoprobes (dCB) have been developed with different outer diameters: 1.1 mm (CB11-S) that can be retracted through the working channel of the bronchoscope, 1.7 mm (CB17) and 2.4 mm (CB24n), respectively. The aim was to evaluate the new cryoprobes with regard to feasibility, specimen area, specimen quality and complications. We compared biopsy samples of the new probes with those obtained by FB and by rCB in an in vivo (porcine) model. A flexible bronchoscope was used to perform biopsy at 4 different locations at the upper and lower lobes of the right and left lung, respectively. The biopsies were taken under fluoroscopic control. The biopsy tool and activation times were allocated randomly. Altogether 204 biopsy procedures were performed. The sample quality of the dCB was superior to that of FB (all p < 0.05) and not significantly different to the rCB sample quality. Mean specimen sample area of all CB was significantly larger compared to FB (p < 0.05). The sample area of the small cryoprobe (CB11-S) was significantly smaller compared to the other CB probes (p < 0.05). No severe bleedings occurred. Pneumothoraces were detected in 3 of the 7 pigs. We conclude that CB with the new single-use instruments are feasible and represent a viable option to improve the diagnostic accuracy of histopathological evaluation compared to FB.

Sections du résumé

BACKGROUND BACKGROUND
Forceps biopsy (FB) is still the most popular procedure for the bronchoscopic sampling of lung tissue. However, it has limitations like inadequate sample size and crush artifacts. Cryobiopsy (CB) has been introduced to obtain bronchoscopic biopsies with improved diagnostic yield compared to FB. Limitations of CB are the need to retract the cryoprobe en bloc with bronchoscope because samples are larger than the working channel and the variations of the freezing power of the reusable CB (rCB). Therefore, 3 new disposable cryoprobes (dCB) have been developed with different outer diameters: 1.1 mm (CB11-S) that can be retracted through the working channel of the bronchoscope, 1.7 mm (CB17) and 2.4 mm (CB24n), respectively.
OBJECTIVES OBJECTIVE
The aim was to evaluate the new cryoprobes with regard to feasibility, specimen area, specimen quality and complications.
METHODS METHODS
We compared biopsy samples of the new probes with those obtained by FB and by rCB in an in vivo (porcine) model. A flexible bronchoscope was used to perform biopsy at 4 different locations at the upper and lower lobes of the right and left lung, respectively. The biopsies were taken under fluoroscopic control. The biopsy tool and activation times were allocated randomly. Altogether 204 biopsy procedures were performed.
RESULTS RESULTS
The sample quality of the dCB was superior to that of FB (all p < 0.05) and not significantly different to the rCB sample quality. Mean specimen sample area of all CB was significantly larger compared to FB (p < 0.05). The sample area of the small cryoprobe (CB11-S) was significantly smaller compared to the other CB probes (p < 0.05). No severe bleedings occurred. Pneumothoraces were detected in 3 of the 7 pigs.
CONCLUSION CONCLUSIONS
We conclude that CB with the new single-use instruments are feasible and represent a viable option to improve the diagnostic accuracy of histopathological evaluation compared to FB.

Identifiants

pubmed: 32101862
pii: 000506017
doi: 10.1159/000506017
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

248-256

Informations de copyright

© 2020 S. Karger AG, Basel.

Auteurs

Jürgen Hetzel (J)

Department of Medical Oncology and Pneumology, University of Tübingen, Tübingen, Germany, juergen.hetzel@med.uni-tuebingen.de.

Walter Linzenbold (W)

Erbe Elektromedizin GmbH, Tübingen, Germany.

Hans Boesmueller (H)

Institute of Pathology and Neuropathology, University of Tübingen, Tübingen, Germany.

Markus Enderle (M)

Erbe Elektromedizin GmbH, Tübingen, Germany.

Venerino Poletti (V)

Department of Diseases of the Thorax, Ospedale GB Morgagni, Forlì, Italy.
Department of Respiratory Diseases and Allergology, Aarhus University Hospital, Aarhus, Denmark.

Articles similaires

Robotic Surgical Procedures Animals Humans Telemedicine Models, Animal

Odour generalisation and detection dog training.

Lyn Caldicott, Thomas W Pike, Helen E Zulch et al.
1.00
Animals Odorants Dogs Generalization, Psychological Smell
Animals TOR Serine-Threonine Kinases Colorectal Neoplasms Colitis Mice
Animals Tail Swine Behavior, Animal Animal Husbandry

Classifications MeSH