Comparison of the Abbott Alinity m and m2000 assays for the quantification of HIV-1, HCV and HBV in clinical samples.


Journal

Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology
ISSN: 1873-5967
Titre abrégé: J Clin Virol
Pays: Netherlands
ID NLM: 9815671

Informations de publication

Date de publication:
05 2020
Historique:
received: 07 10 2019
revised: 09 03 2020
accepted: 09 03 2020
pubmed: 18 3 2020
medline: 29 5 2021
entrez: 18 3 2020
Statut: ppublish

Résumé

Viral load (VL) determination is an essential parameter of the management of patients infected with HIV, HBV or HCV. Many available molecular systems run on a "batch" mode while "random access" systems provide more flexibility. We compared the performance of HIV-1, HCV and HBV quantification assays on the recently developed Abbott Alinity m system to the m2000 RealTime assays. Plasma specimens sent for viral load determination were prospectively tested on m2000 and Alinity m systems, according to manufacturers' instructions. Additional low and high tittered samples were used to assess reproducibility. Assays concordance was evaluated from 180 samples for HIV-1, 122 for HBV, and 92 for HCV. A good correlation and a linear relation over the quantification range was observed for the three markers (r > 0.974). The Alinity m assays yielded higher results with a mean quantification bias of 0.22 log cp/ml for 75 HIV-1, 0.3 log IU/ml for 79 HBV, and 0.2 log for 35 HCV samples, though results were equivalent within an allowable difference of 0.3-0.4 log. Qualitative discordance was observed for 43/180 HIV results, 10/122 HBV and 7/92 HCV and involved undetectable or low-level VL. The Alinity m assays have performance equivalent to m2000. Upon implementation, physicians should be aware of the relative overquantification compared to previous Abbott assays, particularly around clinical decision thresholds. With reduced turnarounds and hands-on times compared to the m2000 system, the Alinity m platform may improve significantly the laboratory workflow efficiency for the benefit of physicians and patients.

Sections du résumé

BACKGROUND
Viral load (VL) determination is an essential parameter of the management of patients infected with HIV, HBV or HCV. Many available molecular systems run on a "batch" mode while "random access" systems provide more flexibility.
OBJECTIVES
We compared the performance of HIV-1, HCV and HBV quantification assays on the recently developed Abbott Alinity m system to the m2000 RealTime assays.
STUDY DESIGN
Plasma specimens sent for viral load determination were prospectively tested on m2000 and Alinity m systems, according to manufacturers' instructions. Additional low and high tittered samples were used to assess reproducibility.
RESULTS
Assays concordance was evaluated from 180 samples for HIV-1, 122 for HBV, and 92 for HCV. A good correlation and a linear relation over the quantification range was observed for the three markers (r > 0.974). The Alinity m assays yielded higher results with a mean quantification bias of 0.22 log cp/ml for 75 HIV-1, 0.3 log IU/ml for 79 HBV, and 0.2 log for 35 HCV samples, though results were equivalent within an allowable difference of 0.3-0.4 log. Qualitative discordance was observed for 43/180 HIV results, 10/122 HBV and 7/92 HCV and involved undetectable or low-level VL.
CONCLUSION
The Alinity m assays have performance equivalent to m2000. Upon implementation, physicians should be aware of the relative overquantification compared to previous Abbott assays, particularly around clinical decision thresholds. With reduced turnarounds and hands-on times compared to the m2000 system, the Alinity m platform may improve significantly the laboratory workflow efficiency for the benefit of physicians and patients.

Identifiants

pubmed: 32179444
pii: S1386-6532(20)30073-1
doi: 10.1016/j.jcv.2020.104331
pii:
doi:

Substances chimiques

RNA, Viral 0
Reagent Kits, Diagnostic 0

Types de publication

Comparative Study Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

104331

Informations de copyright

Copyright © 2020 Elsevier B.V. All rights reserved.

Déclaration de conflit d'intérêts

Declaration of Competing Interest None

Auteurs

Lina Mouna (L)

Virologie, AP-HP, Hôpital Paul Brousse, Villejuif, 94804, France; INSERM U1193, Villejuif, 94804, France.

Coralie Pallier (C)

Virologie, AP-HP, Hôpital Paul Brousse, Villejuif, 94804, France.

Stéphanie Proust (S)

Virologie, AP-HP, Hôpital Paul Brousse, Villejuif, 94804, France.

Corinne Prégermain (C)

Virologie, AP-HP, Hôpital Paul Brousse, Villejuif, 94804, France.

Anne-Marie Roque-Afonso (AM)

Virologie, AP-HP, Hôpital Paul Brousse, Villejuif, 94804, France; INSERM U1193, Villejuif, 94804, France; Univ Paris-Sud Villejuif, 94804, France. Electronic address: anne-marie.roque@aphp.fr.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH