Slice NEQ and system DQE to assess CT imaging performance.
Journal
Physics in medicine and biology
ISSN: 1361-6560
Titre abrégé: Phys Med Biol
Pays: England
ID NLM: 0401220
Informations de publication
Date de publication:
09 06 2020
09 06 2020
Historique:
pubmed:
18
3
2020
medline:
10
10
2020
entrez:
18
3
2020
Statut:
epublish
Résumé
The standard Fourier metrics used for assessing image quality in computed tomography (CT) use a planar (2D) formulation of the noise equivalent quanta (NEQ) without a specific concept to manage the influence of longitudinal resolution variations, thus create a bias in the comparison of image quality for different reconstructed slice thicknesses. For this work, we designed a 'slice NEQ' that takes the longitudinal resolution into account and provides a volumetric indication of the scanner imaging performance. We also developed a specific formulation for the system DQE at the CT isocentre. A cylindrical water phantom with three different inserts for three contrast levels (-100, 120 and 1000 HU) was used for the task-based transfer functions (TTF) and noise power spectra (NPS) measurements. The longitudinal TTF was measured using the point source of the Catphan