Evaluation of the analytical performances of two Raman handheld spectrophotometers for pharmaceutical solid dosage form quantitation.

Accuracy profile method validation Pharmaceutical tablets Quantitative performances Raman handheld spectrophotometer Uniformity of dosage unit

Journal

Talanta
ISSN: 1873-3573
Titre abrégé: Talanta
Pays: Netherlands
ID NLM: 2984816R

Informations de publication

Date de publication:
01 Jul 2020
Historique:
received: 16 12 2019
revised: 27 02 2020
accepted: 28 02 2020
entrez: 13 4 2020
pubmed: 13 4 2020
medline: 13 4 2020
Statut: ppublish

Résumé

This paper addresses the issue of pharmaceutical solid dosage form quantitation using handheld Raman spectrophotometers. The two spectrophotometers used are designed with different technologies: one allows getting a more representative sampling with the Orbital Raster Scanning technology and the other one allows setting acquisition parameters. The goal was to evaluate which technology could provide the best analytical results. Several parameters were optimized to get the lowest prediction error in the end. The main objective of this study was to evaluate if this kind of instrument would be able to identify substandard medicines. For that purpose, two case-study were explored. At first, a full ICH Q2 (R1) compliant validation was performed for moderate Raman scatterer active pharmaceutical ingredient (API) in a specific formulation. It was successfully validated in the ±15% relative total error acceptance limits, with a RMSEP of 0.85% (w/w). Subsequently, it was interesting to evaluate the influence of excipients when the API is a high Raman scatterer. For that purpose, a multi-formulation model was developed and successfully validated with a RMSEP of 2.98% (w/w) in the best case. These two studies showed that thanks to the optimization of acquisition parameters, Raman handheld spectrophotometers methods were validated for two different case-study and could be applied to identify substandard medicines.

Identifiants

pubmed: 32278435
pii: S0039-9140(20)30179-X
doi: 10.1016/j.talanta.2020.120888
pii:
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

120888

Informations de copyright

Copyright © 2020 Elsevier B.V. All rights reserved.

Déclaration de conflit d'intérêts

Declaration of competing interests The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Auteurs

Laureen Coic (L)

University of Liege (ULiege), CIRM, Vibra-Santé Hub, Laboratory of Pharmaceutical Analytical Chemistry, Avenue Hippocrate 15, 4000, Liege, Belgium. Electronic address: laureen.coic@uliege.be.

Pierre-Yves Sacré (PY)

University of Liege (ULiege), CIRM, Vibra-Santé Hub, Laboratory of Pharmaceutical Analytical Chemistry, Avenue Hippocrate 15, 4000, Liege, Belgium.

Amandine Dispas (A)

University of Liege (ULiege), CIRM, Vibra-Santé Hub, Laboratory of Pharmaceutical Analytical Chemistry, Avenue Hippocrate 15, 4000, Liege, Belgium; University of Liege (ULiege), CIRM, MaS-Santé Hub, Laboratory for the Analysis of Medicines, Avenue Hippocrate 15, 4000, Liege, Belgium.

Elodie Dumont (E)

University of Liege (ULiege), CIRM, Vibra-Santé Hub, Laboratory of Pharmaceutical Analytical Chemistry, Avenue Hippocrate 15, 4000, Liege, Belgium.

Julie Horne (J)

University of Liege (ULiege), CIRM, Vibra-Santé Hub, Laboratory of Pharmaceutical Analytical Chemistry, Avenue Hippocrate 15, 4000, Liege, Belgium.

Charlotte De Bleye (C)

University of Liege (ULiege), CIRM, Vibra-Santé Hub, Laboratory of Pharmaceutical Analytical Chemistry, Avenue Hippocrate 15, 4000, Liege, Belgium.

Marianne Fillet (M)

University of Liege (ULiege), CIRM, MaS-Santé Hub, Laboratory for the Analysis of Medicines, Avenue Hippocrate 15, 4000, Liege, Belgium.

Philippe Hubert (P)

University of Liege (ULiege), CIRM, Vibra-Santé Hub, Laboratory of Pharmaceutical Analytical Chemistry, Avenue Hippocrate 15, 4000, Liege, Belgium.

Eric Ziemons (E)

University of Liege (ULiege), CIRM, Vibra-Santé Hub, Laboratory of Pharmaceutical Analytical Chemistry, Avenue Hippocrate 15, 4000, Liege, Belgium.

Classifications MeSH