Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head.
Cardiopulmonary Bypass
Child
Congenital Heart Disease
Drainage
Erythrocytes
Pressure
Vacuum
Journal
Brazilian journal of cardiovascular surgery
ISSN: 1678-9741
Titre abrégé: Braz J Cardiovasc Surg
Pays: Brazil
ID NLM: 101677045
Informations de publication
Date de publication:
01 04 2020
01 04 2020
Historique:
entrez:
6
5
2020
pubmed:
6
5
2020
medline:
21
10
2020
Statut:
epublish
Résumé
To analyze the impact of vacuum-assisted venous drainage (VAVD) on arterial pump flow in a simulated pediatric cardiopulmonary bypass circuit utilizing a centrifugal pump (CP) with an external arterial filter. The simulation circuit consisted of a Quadrox-I Pediatric oxygenator, a Rotaflow CP (Maquet Cardiopulmonary AG, Rastatt, Germany), and a custom pediatric tubing set primed with Lactated Ringer's solution and packed red blood cells. Venous line pressure, reservoir pressure, and arterial flow were measured with VAVD turned off to record baseline values. Four other conditions were tested with progressively higher vacuum pressures (-20, -40, -60, and -80 mmHg) applied to the baseline cardiotomy pressure. An arterial filter was placed into the circuit and arterial flow was measured with the purge line in both open and closed positions. These trials were repeated at set arterial flow rates of 1500, 2000, and 2500 mL/min. The use of progressively higher vacuum caused a reduction in effective arterial flow from 1490±0.00 to 590±0.00, from 2020±0.01 to 1220±0.00, and from 2490±0.0 to 1830±0.01 mL/min. Effective forward flow decreased with increased levels of VAVD. The use of VAVD reduces arterial flow when a CP is used as the main arterial pump. The reduction in the forward arterial flow increases as the vacuum level increases. The loss of forward flow is further reduced when the arterial filter purge line is kept in the recommended open position.An independent flow probe is essential to monitor pump flow during cardiopulmonary bypass.
Identifiants
pubmed: 32369291
doi: 10.21470/1678-9741-2019-0311
pmc: PMC7199987
doi:
Types de publication
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM
Pagination
134-140Références
Anesth Analg. 2009 May;108(5):1394-417
pubmed: 19372313
JNMA J Nepal Med Assoc. 2014 Oct-Dec;52(196):960-6
pubmed: 26982892
Eur J Cardiothorac Surg. 2007 Jun;31(6):1044-51
pubmed: 17336540
Rev Bras Cir Cardiovasc. 2014 Apr-Jun;29(2):266-71
pubmed: 25140478
J Thorac Cardiovasc Surg. 2000 Nov;120(5):856-62
pubmed: 11044310
Ann Thorac Surg. 1999 Oct;68(4):1285-9
pubmed: 10543494
J Extra Corpor Technol. 2003 Sep;35(3):207-11
pubmed: 14653422
Artif Organs. 2012 Aug;36(8):668-76
pubmed: 22804106
Perfusion. 2020 Jan;35(1):26-31
pubmed: 31146643
J Extra Corpor Technol. 2006 Sep;38(3):220-9
pubmed: 17089508
Artif Organs. 2020 Jan;44(1):28-39
pubmed: 30512218
ASAIO Trans. 1991 Jul-Sep;37(3):M487-9
pubmed: 1836339
Braz J Cardiovasc Surg. 2018 May-Jun;33(3):224-232
pubmed: 30043914
J Extra Corpor Technol. 2015 Sep;47(3):160-6
pubmed: 26543250
Artif Organs. 1998 Apr;22(4):326-36
pubmed: 9555964
J Extra Corpor Technol. 2008 Dec;40(4):249-56
pubmed: 19192754
Braz J Cardiovasc Surg. 2019 Mar-Apr;34(2):239-260
pubmed: 30916138