Dual-fiber point diffraction interferometer to measure the wavefront aberration of an imaging system.
Journal
Applied optics
ISSN: 1539-4522
Titre abrégé: Appl Opt
Pays: United States
ID NLM: 0247660
Informations de publication
Date de publication:
01 Apr 2020
01 Apr 2020
Historique:
entrez:
14
5
2020
pubmed:
14
5
2020
medline:
14
5
2020
Statut:
ppublish
Résumé
A dual-fiber point diffraction interferometer is built to measure the wavefront aberration of an imaging system with high precision. The optical intensity of the test light and reference light are controlled independently, so that high interference contrast can be obtained. The interferometer has several advantages: high precision, a flexible structure, and a quasi-common optical path. System errors of the interferometer that influence the measurement accuracy are considered and calibrated. In the experiments, a projection lens with numerical aperture of 0.3 is measured. The measured wavefront error is 1.5 nm rms, and the measurement repeatability is 35 pm rms.
Identifiants
pubmed: 32400590
pii: 429424
doi: 10.1364/AO.387540
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM