Combination system in advanced image processing for improving image contrast and a conventional row-ratio grid for an indirect flat-panel detector system: An experimental study.
Anti-scatter grid
Flat-panel detector
Image quality
Removal processing software for scattered radiation
Journal
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)
ISSN: 1724-191X
Titre abrégé: Phys Med
Pays: Italy
ID NLM: 9302888
Informations de publication
Date de publication:
Jun 2020
Jun 2020
Historique:
received:
31
01
2020
revised:
03
04
2020
accepted:
21
04
2020
pubmed:
16
5
2020
medline:
7
4
2021
entrez:
16
5
2020
Statut:
ppublish
Résumé
Conventional grids with high grid ratios are not ideal for use in bedside radiography because of the difficulty in maintaining the required alignment. To address this issue, the potential usefulness of a combination system that employs removal processing software for scattered radiation and a conventional grid with a low grid ratio (3:1) for an indirect-conversion-type flat-panel detector system was evaluated by measuring image quality and observer performance. The hypothetical grid ratios for the software were 2:1, 3:1, 6:1, 8:1, and 10:1. The scatter fraction of the combination system was lower than that of the software alone. Significant improvement was observed in the effect of scattered radiation removal up to a hypothetical software grid ratio of 6:1. However, the Wiener spectrum increased (radiographic noise degraded) with an increase in the hypothetical grid ratio. The contrast ratios of the combination system were improved compared to those of the software alone for anthropomorphic chest radiographs. An observer test was also conducted using the contrast-detail phantom. The combination system indicated higher low-contrast detectability compared to the software alone, although there were no statistical differences between the hypothetical grid ratios of 6:1, 8:1, and 10:1 in all combinations of the software alone and the combination system. We concluded that a combination system with software that uses a hypothetical grid ratio of 6:1 or more and a 3:1 conventional grid would be more useful for reducing the scattered radiation component compared to the software alone with a hypothetical higher grid ratio for thicker objects.
Identifiants
pubmed: 32413514
pii: S1120-1797(20)30098-3
doi: 10.1016/j.ejmp.2020.04.017
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
40-46Informations de copyright
Copyright © 2020 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.