Dynamic K-edge Subtraction Fluoroscopy at a Compact Inverse-Compton Synchrotron X-ray Source.


Journal

Scientific reports
ISSN: 2045-2322
Titre abrégé: Sci Rep
Pays: England
ID NLM: 101563288

Informations de publication

Date de publication:
15 06 2020
Historique:
received: 05 03 2020
accepted: 20 05 2020
entrez: 17 6 2020
pubmed: 17 6 2020
medline: 15 12 2020
Statut: epublish

Résumé

X-ray fluoroscopy is a commonly applied diagnostic tool for morphological and functional evaluation of the intestine in clinical routine. Acquisition of repetitive X-ray images following oral or rectal application of iodine contrast agent visualizes the time dependent distribution of the contrast medium, and helps to detect for example leakages, tumors or functional disorders. However, movements of the intestine and air trapped inside usually prevent temporal subtraction imaging to be applied to fluoroscopy of the gastrointestinal tract. K-edge subtraction (KES) imaging would enable subtraction fluoroscopy because it allows for imaging of moving organs with little artefacts. Although KES imaging is a well established technique at synchrotron sources, this imaging method is not applied in clinical routine as it relies on brilliant synchrotron radiation. Recently emerging compact synchrotron X-ray sources could provide a quasi-monochromatic, high-flux X-ray beam and allow for the application of KES in a laboratory environment. Here, we present a filter-based dynamic KES approach at the Munich Compact Light Source (MuCLS), the first user-dedicated installation of a compact synchrotron X-ray source worldwide. Compared to conventional temporal subtraction X-ray radiography, our approach increases the contrast while reducing the generated image artefacts.

Identifiants

pubmed: 32541788
doi: 10.1038/s41598-020-66414-x
pii: 10.1038/s41598-020-66414-x
pmc: PMC7295988
doi:

Substances chimiques

Contrast Media 0

Types de publication

Journal Article Research Support, Non-U.S. Gov't

Langues

eng

Sous-ensembles de citation

IM

Pagination

9612

Références

Meijering, E. H. W., Zuiderveld, K. J. & Viergever, M. A. X. A. Image Registration for Digital Subtraction Angiography. Int. J. Comput. Vis. 31, 227–246 (1999).
doi: 10.1023/A:1008074100927
Tobis, J. M., Nalcioglu, O. & Henry, W. L. Digital Subtraction Angiography. Chest 84, 63–86 (1983).
doi: 10.1378/chest.84.1.68
Yamamoto, M. et al. Development of Digital Subtraction Angiography for Coronary Artery. J. Digit. Imaging 22, 319–325 (2009).
doi: 10.1007/s10278-008-9108-1
Winklhofer, S. et al. Reduction of peristalsis-related gastrointestinal streak artifacts with dual-energy CT: a patient and phantom study. Abdom Radiol (NY) 41, 1456–1465 (2016).
doi: 10.1007/s00261-016-0702-2
Jacobson, B. Dichromatic Absorption Radiography. Dichromography. Acta Radiol. 39, 437–452 (1953).
doi: 10.3109/00016925309136730
Sarnelli, A. et al. K-edge digital subtraction imaging based on a dichromatic and compact x-ray source. Phys. Medicine Biol. 49, 3291–3305 (2004).
doi: 10.1088/0031-9155/49/14/019
Arhatari, B. D., Gureyev, T. E. & Abbey, B. Elemental Contrast X-ray Tomography Using Ross Filter Pairs with a Polychromatic Laboratory Source. Sci. Reports 7, 218 (2017).
Roessl, E. & Proksa, R. K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors. Phys. Medicine Biol. 52, 4679–4696 (2007).
doi: 10.1088/0031-9155/52/15/020
Schlomka, J. P. et al. Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography. Phys. Medicine Biol. 53, 4031–4047 (2008).
doi: 10.1088/0031-9155/53/15/002
Shikhaliev, P. M. Computed tomography with energy-resolved detection: a feasibility study. Phys. Medicine Biol. 53, 1475–1495 (2008).
doi: 10.1088/0031-9155/53/5/020
Vincenzi, A. et al. Energy characterization of Pixirad-1 photon counting detector system. J. Instrumentation 10, C04010 (2015).
doi: 10.1088/1748-0221/10/04/C04010
Pani, S. et al. Optimization of K-edge subtraction imaging using a pixellated spectroscopic detector. IEEE Nucl. Sci. Symp. Conf. Rec. 3063–3066, https://doi.org/10.1109/NSSMIC.2012.6551700 (2012).
Egan, C. K. et al. 3D chemical imaging in the laboratory by hyperspectral X-ray computed tomography. Sci. Reports 5, 15979 (2015).
Rubenstein, E. et al. Synchrotron Radiation for Transvenous Coronary Angiography. Trans. Am. Clin. Clim. Assoc. 97, 27–31 (1986).
Dix, W.-R. Intravenous coronary angiography with synchrotron radiation. Prog. Biophys. Mol. Biol. 63, 159–191 (1995).
doi: 10.1016/0079-6107(95)00003-6
Elleaume, H. et al. First human transvenous coronary angiography at the European Synchrotron Radiation Facility. Phys. Med. Biol. 45, L39–L43 (2000).
doi: 10.1088/0031-9155/45/9/102
Bertrand, B. et al. Comparison of synchrotron radiation angiography with conventional angiography for the diagnosis of in-stent restenosis after percutaneous transluminal coronary angioplasty. Eur. Hear. J. 26, 1284–1291 (2005).
doi: 10.1093/eurheartj/ehi165
Dilmanian, F. A. et al. Single- and dual-energy CT with monochromatic synchrotron x-rays. Phys. Medicine Biol. 42, 371–387 (1997).
doi: 10.1088/0031-9155/42/2/009
Schültke, E. et al. Dual energy CT at the synchrotron: A piglet model for neurovascular research. Eur. J. Radiol. 79, 323–327 (2011).
doi: 10.1016/j.ejrad.2010.07.002
Cooper, D. M. et al. Three dimensional mapping of strontium in bone by dual energy K-edge subtraction imaging. Phys. Medicine Biol. 57, 5777–5786 (2012).
doi: 10.1088/0031-9155/57/18/5777
Strengell, S. et al. Radiation dose and image quality in K-edge subtraction computed tomography of lung in vivo. J. Synchrotron Radiat. 21, 1305–1313 (2014).
doi: 10.1107/S160057751401697X
Thomlinson, W., Elleaume, H., Porra, L. & Suortti, P. K-edge subtraction synchrotron X-ray imaging in bio-medical research. Phys. Medica 49, 58–76 (2018).
doi: 10.1016/j.ejmp.2018.04.389
Schültke, E. et al. The potential for neurovascular intravenous angiography using K-edge digital subtraction angiography. Nucl. Instruments Methods Phys. Res. A 548, 84–87 (2005).
doi: 10.1016/j.nima.2005.03.071
Elleaume, H., Charvet, A. M., Corde, S., Estève, F. & Le Bas, J. F. Performance of computed tomography for contrast agent concentration measurements with monochromatic x-ray beams: comparison of K-edge versus temporal subtraction. Phys. Med. Biol. 47, 3369–3385 (2002).
doi: 10.1088/0031-9155/47/18/307
Houk, T. L. et al. Real-Time Digital K-Edge Subtraction Fluoroscopy. Investig. Radiol. 14, 270–278 (1979).
doi: 10.1097/00004424-197907000-00002
Huang, Z. & Ruth, R. D. Laser-Electron Storage Ring. Phys. Rev. Lett. 80, 976–979 (1998).
doi: 10.1103/PhysRevLett.80.976
Jacquet, M. High intensity compact Compton X-ray sources: Challenges and potential of applications. Nucl. Instruments Methods Phys. Res. Sect. B 331, 1–5 (2014).
doi: 10.1016/j.nimb.2013.10.078
Bacci, A. et al. Status of the STAR Project. Proc. IPAC2016, Busan, Korea TUPOW004 (2016).
Eggl, E. et al. The Munich Compact Light Source: initial performance measures. J. Synchrotron Radiat. 23, 91–100 (2016).
doi: 10.1107/S160057751600967X
Eggl, E. et al. Mono-Energy Coronary Angiography with a Compact Synchtrotron Source. Sci. Reports 7, 42211, (2017). 1506.01997.
Kulpe, S. et al. K-edge subtraction imaging for coronary angiography with a compact synchrotron X-ray source. PloS ONE 13, e0208446 (2018).
doi: 10.1371/journal.pone.0208446
Kulpe, S. et al. K-edge Subtraction Computed Tomography with a Compact Synchrotron X-ray Source. Sci. Reports 9, 13332 (2019).
Thomas, J. V. et al. Gadoxetate Disodium enhanced spectral dual-energy CT for evaluation of cholangiocarcinoma: Preliminary data. Annals Medicine Surg. 6, 17–22 (2016).
doi: 10.1016/j.amsu.2016.01.001
van Hamersvelt, R. W. et al. Feasibility and accuracy of dual-layer spectral detector computed tomography for quantification of gadolinium: a phantom study. Eur. Radiol. 27, 3677–3686 (2017).
doi: 10.1007/s00330-017-4737-8
Sam, A. D. et al. Safety of gadolinium contrast angiography in patients with chronic renal insufficiency. J. Vasc. Surg. 38, 313–318 (2003).
doi: 10.1016/S0741-5214(03)00315-X
Sayin, T., Turhan, S., Akyürek, O. & Kilickap, M. Gadolinium:nonionic contrast media (1:1) coronary angiography in patients with impaired renal function. Angiology 58, 561–4 (2007).
doi: 10.1177/0003319707303640
Variola, A. The ThomX project. Proc. IPAC2011, San Sebastian, Spain WEOAA01 (2011).
Faillace, L. et al. Status of compact inverse Compton sources in Italy: BriXS and STAR. Proc. SPIE Opt. Photonics 2019 1111005, (September 2019).
Achterhold, K. et al. Monochromatic computed tomography with a compact laser-driven X-ray source. Sci. Reports 3, 1313 (2013).
Loewen, R. J. A Compact Light Source: Design and Technical Feasibility Study of a Laser-Electron Storage Ring X-ray Source. Ph.D. thesis, Standford University, https://doi.org/10.2172/826755 (2003).
Günther, B. et al. The Munich Compact Light Source: Flux Doubling and Source Position Stabilization At a Compact Inverse-Compton Synchrotron X-ray Source. Microsc. Microanal. 24, 312–313 (2018).
Umetani, K. et al. Iodine K-edge dual-energy imaging for subtraction angiography using synchrotron radiation and a 2-dimensional detector. Nucl. Instruments Methods Phys. Res. A 301, 579–588 (1991).
doi: 10.1016/0168-9002(91)90026-M
Umetani, K. et al. Iodine filter imaging system for subtraction angiography using synchrotron radiation. Nucl. Instruments Methods Phys. Res. A 335, 569–579 (1993).
doi: 10.1016/0168-9002(93)91245-I
Murrie, R. P., Paganin, D. M., Fouras, A. & Morgan, K. S. Phase contrast x-ray velocimetry of small animal lungs: optimising imaging rates. Biomed. Opt. Express 7, 79–92 (2016).
doi: 10.1364/BOE.7.000079
Gradl, R. et al. In vivo Dynamic Phase-Contrast X-ray Imaging using a Compact Light Source. Sci. Reports 8, 6788 (2018).

Auteurs

Stephanie Kulpe (S)

Chair of Biomedical Physics, Department of Physics, Technical University of Munich, James-Franck-Str. 1, 85748, Garching, Germany. stephanie.kulpe@tum.de.
Munich School of BioEngineering, Technical University of Munich, Boltzmannstr. 11, 85748, Garching, Germany. stephanie.kulpe@tum.de.

Martin Dierolf (M)

Chair of Biomedical Physics, Department of Physics, Technical University of Munich, James-Franck-Str. 1, 85748, Garching, Germany.
Munich School of BioEngineering, Technical University of Munich, Boltzmannstr. 11, 85748, Garching, Germany.

Benedikt Günther (B)

Chair of Biomedical Physics, Department of Physics, Technical University of Munich, James-Franck-Str. 1, 85748, Garching, Germany.
Munich School of BioEngineering, Technical University of Munich, Boltzmannstr. 11, 85748, Garching, Germany.

Johannes Brantl (J)

Chair of Biomedical Physics, Department of Physics, Technical University of Munich, James-Franck-Str. 1, 85748, Garching, Germany.
Munich School of BioEngineering, Technical University of Munich, Boltzmannstr. 11, 85748, Garching, Germany.

Madleen Busse (M)

Chair of Biomedical Physics, Department of Physics, Technical University of Munich, James-Franck-Str. 1, 85748, Garching, Germany.
Munich School of BioEngineering, Technical University of Munich, Boltzmannstr. 11, 85748, Garching, Germany.

Klaus Achterhold (K)

Chair of Biomedical Physics, Department of Physics, Technical University of Munich, James-Franck-Str. 1, 85748, Garching, Germany.
Munich School of BioEngineering, Technical University of Munich, Boltzmannstr. 11, 85748, Garching, Germany.

Bernhard Gleich (B)

Munich School of BioEngineering, Technical University of Munich, Boltzmannstr. 11, 85748, Garching, Germany.

Franz Pfeiffer (F)

Chair of Biomedical Physics, Department of Physics, Technical University of Munich, James-Franck-Str. 1, 85748, Garching, Germany.
Munich School of BioEngineering, Technical University of Munich, Boltzmannstr. 11, 85748, Garching, Germany.
Department of Diagnostic and Interventional Radiology, Munich School of Medicine and Klinikum rechts der Isar, Ismaniger Str. 22, 81675, Munich, Germany.

Daniela Pfeiffer (D)

Department of Diagnostic and Interventional Radiology, Munich School of Medicine and Klinikum rechts der Isar, Ismaniger Str. 22, 81675, Munich, Germany.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH