Spatial Response Identification for Flexible and Accurate Ultrasound Transducer Calibration and its Application to Brain Imaging.


Journal

IEEE transactions on ultrasonics, ferroelectrics, and frequency control
ISSN: 1525-8955
Titre abrégé: IEEE Trans Ultrason Ferroelectr Freq Control
Pays: United States
ID NLM: 9882735

Informations de publication

Date de publication:
01 2021
Historique:
pubmed: 11 8 2020
medline: 26 10 2021
entrez: 11 8 2020
Statut: ppublish

Résumé

Accurate wave-equation modeling is becoming increasingly important in modern imaging and therapeutic ultrasound methodologies, such as ultrasound computed tomography, optoacoustic tomography, or high-intensity-focused ultrasound. All of them rely on the ability to accurately model the physics of wave propagation, including accurate characterization of the ultrasound transducers, the physical devices that are responsible for generating and recording ultrasound energy. However, existing methods fail to characterize the transducer response with the accuracy required to fully exploit the capabilities of these emerging imaging and therapeutic techniques. Consequently, we have designed a new algorithm for ultrasound transducer calibration and modeling: spatial response identification (SRI). This method introduces a parameterization of the ultrasound transducer and provides a method to calibrate the transducer model using experimental data, based on a formulation of the problem that is completely independent of the discretization chosen for the transducer or the number of parameters used. The proposed technique models the transducer as a linear time-invariant system that is spatially heterogeneous, and identifies the model parameters that are best at explaining the experimental data while honoring the full wave equation. SRI generates a model that can accommodate the complex, heterogeneous spatial response seen experimentally for ultrasound transducers. Experimental results show that SRI outperforms standard methods both in transmission and reception modes. Finally, numerical experiments using full-waveform inversion demonstrate that existing transducer-modeling approaches are insufficient to produce successful reconstructions of the human brain, whereas errors in our SRI algorithm are sufficiently small to allow accurate image reconstructions.

Identifiants

pubmed: 32776878
doi: 10.1109/TUFFC.2020.3015583
doi:

Types de publication

Journal Article Research Support, Non-U.S. Gov't

Langues

eng

Sous-ensembles de citation

IM

Pagination

143-153

Auteurs

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH