Use, reuse or discard: quantitatively defined variance in N95 respirator integrity following vaporized hydrogen peroxide decontamination during the COVID-19 pandemic.
Journal
medRxiv : the preprint server for health sciences
Titre abrégé: medRxiv
Pays: United States
ID NLM: 101767986
Informations de publication
Date de publication:
21 Aug 2020
21 Aug 2020
Historique:
entrez:
26
8
2020
pubmed:
26
8
2020
medline:
26
8
2020
Statut:
epublish
Résumé
COVID-19 has stretched the ability of many institutions to supply needed personal protective equipment, especially N95 respirators. N95 decontamination and reuse programs provide one potential solution to this problem. Unfortunately, a comprehensive evaluation of the effects of decontamination on the integrity of various N95 models using a quantitative fit test (QTFT) approach is lacking. 1) To investigate the effects of up to eight rounds of vaporized H2O2 (VHP) decontamination on the integrity of N95 respirators currently in use in a hospital setting. 2) To examine if N95 respirators worn by one user can adapt to the face shape of a second user with no compromise of integrity following VHP decontamination. The PortaCount Pro+ Respirator Fit Tester Model 8038 was used to quantitatively define the integrity, measured by fit, of N95 respirators following decontamination with VHP. There was an observable downward trend in the integrity of Halyard Fluidshield 46727 N95 respirators throughout eight cycles of decontamination with VHP. The integrity of 3M 1870 N95 respirators was significantly reduced after the respirator was worn, decontaminated with VHP, and then quantitatively fit tested on a second user. Furthermore, we uncovered inconsistencies between qualitative fit test and QTFT results that may have strong implications on the fit testing method used by institutions. Our data revealed variability in the integrity of different N95 models after VHP decontamination and exposed potential limitations of N95 decontamination and reuse programs.
Sections du résumé
BACKGROUND
BACKGROUND
COVID-19 has stretched the ability of many institutions to supply needed personal protective equipment, especially N95 respirators. N95 decontamination and reuse programs provide one potential solution to this problem. Unfortunately, a comprehensive evaluation of the effects of decontamination on the integrity of various N95 models using a quantitative fit test (QTFT) approach is lacking.
AIMS
OBJECTIVE
1) To investigate the effects of up to eight rounds of vaporized H2O2 (VHP) decontamination on the integrity of N95 respirators currently in use in a hospital setting. 2) To examine if N95 respirators worn by one user can adapt to the face shape of a second user with no compromise of integrity following VHP decontamination.
METHODS
METHODS
The PortaCount Pro+ Respirator Fit Tester Model 8038 was used to quantitatively define the integrity, measured by fit, of N95 respirators following decontamination with VHP.
FINDINGS
RESULTS
There was an observable downward trend in the integrity of Halyard Fluidshield 46727 N95 respirators throughout eight cycles of decontamination with VHP. The integrity of 3M 1870 N95 respirators was significantly reduced after the respirator was worn, decontaminated with VHP, and then quantitatively fit tested on a second user. Furthermore, we uncovered inconsistencies between qualitative fit test and QTFT results that may have strong implications on the fit testing method used by institutions.
CONCLUSIONS
CONCLUSIONS
Our data revealed variability in the integrity of different N95 models after VHP decontamination and exposed potential limitations of N95 decontamination and reuse programs.
Identifiants
pubmed: 32839780
doi: 10.1101/2020.08.18.20177071
pmc: PMC7444293
pii:
doi:
Types de publication
Preprint
Langues
eng
Commentaires et corrections
Type : UpdateIn