Assignment of international normalized ratio to frozen and freeze-dried pooled plasmas.


Journal

Clinical chemistry and laboratory medicine
ISSN: 1437-4331
Titre abrégé: Clin Chem Lab Med
Pays: Germany
ID NLM: 9806306

Informations de publication

Date de publication:
26 11 2020
Historique:
received: 23 12 2019
accepted: 29 03 2020
pubmed: 26 8 2020
medline: 23 6 2021
entrez: 26 8 2020
Statut: ppublish

Résumé

Objectives Frozen and freeze-dried plasmas may be used for local prothrombin time system calibration, for direct international normalized ratio (INR) determination, and for quality assessment. The purpose of the present study was to evaluate the usefulness of INRs assigned with various types of thromboplastins to frozen and freeze-dried pooled plasmas obtained from patients treated with vitamin K antagonists. Methods INRs were calculated according to the international sensitivity index (ISI) model using various thromboplastins and instruments, i.e. International Standards for thromboplastin as well as six commercial reagents prepared from rabbit and bovine brain, and recombinant human tissue factor. The uncertainty of the INRs was assessed using the standard deviations of clotting times and ISI values. Commutability of the plasmas was assessed according to the approved Clinical and Laboratory Standards Institute (CLSI) Guideline EP30-A. Validation of a set of six frozen plasma pools for direct INR determination was performed according to the Subcommittee on Control of Anticoagulation of the Scientific and Standardization Committee of the International Society on Thrombosis and Haemostasis (SSC/ISTH) guidelines. Results For all frozen and freeze-dried plasmas, the INRs calculated with bovine thromboplastin Thrombotest were lower than the INRs assigned with other thromboplastins. With a few exceptions, the frozen and freeze-dried pooled plasmas were commutable. When the set of six frozen plasma pools was used for local calibration, the analytical bias of the INR was less than ±10% for all commercial reagents except Thrombotest. Conclusions Processing of fresh plasmas to prepare pooled frozen plasmas and freeze-dried plasmas may lead to different INR assignments depending on the thromboplastin used. Despite minor INR differences, a set of six frozen plasma pools could be used for local calibration by direct INR determination.

Identifiants

pubmed: 32840098
doi: 10.1515/cclm-2019-1321
pii: cclm-2019-1321
doi:

Substances chimiques

Prothrombin 9001-26-7

Types de publication

Journal Article Research Support, Non-U.S. Gov't

Langues

eng

Sous-ensembles de citation

IM

Pagination

2089-2097

Références

Van den Besselaar AM, Barrowcliffe TW, Houbouyan-Reveillard LL, Jespersen J, Johnston M, Poller L, et al. On behalf of thesubcommittee on control of anticoagulation of the Scientific and Standardization Committee of the ISTH. Guidelines on preparation, certification, and use of certified plasmas for ISI calibration and INR determination. J Thromb Haemost 2004;2:1946–53.
Meijer P, Kynde K, van den Besselaar AM, van Blerk M, Woods TA. International normalized ratio (INR) testing in Europe: between-laboratory comparability of test results obtained by Quick and Owren reagents. Clin Chem Lab Med 2018;56:1698–703.
Fattorini A, Pattarini E, Viganò S, Crippa L, D’Angelo A. Use of INR calibrator plasmas in the routine coagulation laboratory: a study of two thromboplastin reagents. Thromb Res 2012;130:e158–62.
Van den Besselaar AM. Artificially depleted plasmas are not necessarily commutable with native patient plasmas for International Sensitivity Index calibration and International Normalized Ratio derivation. J Thromb Haemost 2012;10:303–5.
Poller L, Keown M, Ibrahim S, van den Besselaar AM, Roberts C, Stevenson K, et al. Comparison of local International Sensitivity Index calibration and ‘Direct INR’ methods in correction of locally reported International Normalized Ratios: an international study. J Thromb Haemost 2007;5:1002–9.
Van den Besselaar AM, Houbouyan-Reveillard LL, Aillaud MF, Denson KW, Droullé C, Johnston M, et al. Multicenter evaluation of lyophilized and deep-frozen plasmas for assignment of the International Normalized Ratio. Thromb Haemost 1999;82:1451–5.
Clinical and Laboratory Standards Institute. Characterization and qualification of commutable reference materials for laboratory medicine; Approved Guideline 2010, CLSI document EP30-A (formerly C53-A). Wayne, PA: CLSI.
Stavelin A, Rikshelm BO, Christensen NG, Sandberg S. The importance of reagent lot registration in external quality assurance/proficiency testing schemes. Clin Chem 2016;62:708–15.
Van den Besselaar AM, Abdoel CF, van Rijn CJ, van der Meer FJ, Cobbaert CM. Accuracy assessment of consecutive test strip lots for whole blood INR point-of-care instruments: clarifying the role of frozen plasmas. Clin Chem Lab Med 2019;57:1349–57.
Owren PA. Thrombotest. A new method for controlling anticoagulant therapy. Lancet 1959;2:754–8.
Laharrague P, Carrières J, Cambus JP, Boneu B, Biermé R. Évaluation du thrombotest d’Owren dans la surveillance des traitements par antivitamine K. Ann Biol clin 1983;41:123–7.
Morrison M, Caldwell A, McQuaker G, Fitzsimons EJ. Discrepant INR values: a comparison between Manchester and Thrombotest reagents using capillary and venous samples. Clin Lab Haemat 1989;11:393–8.
Van den Besselaar AM. Use of lyophilized calibration plasmas and control blood for international normalized ratio calculation in external quality assessment of the prothrombin time. Am J Clin Pathol 1994;102:123–7
Poller L, Barrowcliffe TW, van den Besselaar AM, Jespersen J, Tripodi A, Houghton D. The European Concerted Action on Anticoagulation (ECAA) evaluation of a set of lyophilized normal plasmas to establish the normal prothrombin time for coagulometer systems. Thromb Haemost 1998;79:122–8.
Uetsuka Y, Sakuragawa N, Wada H, Aizawa Y, Aosaki M. ISI/INR system in Japan: experience from simultaneous measurement of the same plasma at four different laboratories. Semin Thromb Hemost 2000;26:109–12
Van den Besselaar AM, Tripodi A. Is there a need for replacement of the International Reference Preparation for thromboplastin, bovine, combined (OBT/79)? J Thromb Haemost 2005;3:2365–6.
WHO Expert Committee on Biological Standardization. Guidelines for thromboplastins and plasmas used to control oral anticoagulant therapy with vitamin K antagonists. WHO Technical Report Series No. 979, 2013:271–316.
Tripodi A, Chantarangkul V, van den Besselaar AM, Witteveen E, Hubbard AR. International collaborative study for the calibration of a proposed international standard for thromboplastin, human, plain. J Thromb Haemost 2010;8:2066–8.
Chantarangkul V, van den Besselaar AM, Witteveen E, Tripodi A. International collaborative study for the calibration of a proposed international standard for thromboplastin, rabbit, plain. J Thromb Haemost 2006;4:1339–45.
Van den Besselaar AM, Bertina RM. Multi-center study of thromboplastin calibration precision – influence of reagent species, composition, and international sensitivity index (ISI). Thromb Haemost 1993;69:35–40.
Van den Besselaar AM, Lewis SM, Mannucci PM, Poller L. Status of present and candidate international reference preparations (IRP) of thromboplastin for the prothrombin time. Thromb Haemost 1993;69:85.
Van den Besselaar AM, Witteveen E, van der Meer FJ. Uncertainty of International Sensitivity Index and International Normalized Ratio. J Thromb Haemost 2013;11:1615–7.
Van den Besselaar AM, Chantarangkul V, Angeloni F, Binder NB, Byrne M, Dauer R, et al. International collaborative study for the calibration of proposed International Standards for thromboplastin, rabbit, plain, and for thromboplastin, recombinant, human, plain. J Thromb Haemost 2018;16:142–9.
Van den Besselaar AM. Comparison of lyophilized plasmas with fresh plasmas for calibration of thromboplastin reagents in oral anticoagulant control. Br J Haematol 1996;93:437–44.
Van den Besselaar AM, van Halem-Visser LP, Loeliger EA. The use of evacuated tubes for blood collection in oral anticoagulant control. Thromb Haemost 1983;50:676–7.
Poggio M, Tripodi A, Mariani G, Mannucci PM. Factor VII clotting assay: influence of different thromboplastins and factor VII-deficient plasmas. CISMEL Study Group. Thromb Haemost 1991;65:160–4.
Roberts PS, Hughes HN, Fleming PB. The effects of hepes buffer on clotting tests, assay of factors V and VIII and on the hydrolysis of esters by thrombin and thrombokinase. Thromb Haemost 1976;35:202–10.
Poller L, Keown M, Shepherd SA, Shiach CR, Tabeart S. The effects of freeze drying and freeze drying additives on the prothrombin time and the international sensitivity index. J Clin Pathol 1999;52:744–8.
Young IS. The enduring importance and challenge of commutability. Clin Chem 2018;64:421–3.
Miller WG, Schimmel H, Rej R, Greenberg N, Ceriotti F, Burns C, et al. IFCC Working Group Recommendations for Assessing Commutability Part 1: General experimental design. Clin Chem 2018;64:447–54.
Nilsson G, Budd JR, Greenberg N, Delatour V, Rej R, Panteghini M, et al. IFCC Working Group Recommendations for Assessing Commutability Part 2: Using the difference in bias between a reference material and clinical samples. Clin Chem 2018;64:455–64.
Budd JR, Weykamp C, Rej R, MacKenzie F, Ceriotti F, Greenberg N, et al. IFCC Working Group Recommendations for Assessing Commutability Part 3: Using the calibration effectiveness of a reference material. Clin Chem 2018;64:465–74.
Van den Besselaar AM. Adjustment of ISI calibration of Innovin. Ned Tijdschr Klin Chem Labgeneesk 2007;32:224–7.

Auteurs

Antonius M H P van den Besselaar (AMHP)

Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, P.O. Box 9600, 2300 RC Leiden, The Netherlands.
Department of Thrombosis and Hemostasis, Leiden University Medical Center, Leiden, The Netherlands, Phone: +31 71 526 1942.

Christa M Cobbaert (CM)

Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, The Netherlands.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH