Integrating and Evaluating the Data Quality and Utility of Smart Pump Information in Detecting Medication Administration Errors: Evaluation Study.
concordance
electronic health records
medication administration errors
smart infusion pumps
Journal
JMIR medical informatics
ISSN: 2291-9694
Titre abrégé: JMIR Med Inform
Pays: Canada
ID NLM: 101645109
Informations de publication
Date de publication:
02 Sep 2020
02 Sep 2020
Historique:
received:
01
05
2020
accepted:
03
07
2020
revised:
01
07
2020
entrez:
3
9
2020
pubmed:
3
9
2020
medline:
3
9
2020
Statut:
epublish
Résumé
At present, electronic health records (EHRs) are the central focus of clinical informatics given their role as the primary source of clinical data. Despite their granularity, the EHR data heavily rely on manual input and are prone to human errors. Many other sources of data exist in the clinical setting, including digital medical devices such as smart infusion pumps. When incorporated with prescribing data from EHRs, smart pump records (SPRs) are capable of shedding light on actions that take place during the medication use process. However, harmoniz-ing the 2 sources is hindered by multiple technical challenges, and the data quality and utility of SPRs have not been fully realized. This study aims to evaluate the quality and utility of SPRs incorporated with EHR data in detecting medication administration errors. Our overarching hypothesis is that SPRs would contribute unique information in the med-ication use process, enabling more comprehensive detection of discrepancies and potential errors in medication administration. We evaluated the medication use process of 9 high-risk medications for patients admitted to the neonatal inten-sive care unit during a 1-year period. An automated algorithm was developed to align SPRs with their medica-tion orders in the EHRs using patient ID, medication name, and timestamp. The aligned data were manually re-viewed by a clinical research coordinator and 2 pediatric physicians to identify discrepancies in medication ad-ministration. The data quality of SPRs was assessed with the proportion of information that was linked to valid EHR orders. To evaluate their utility, we compared the frequency and severity of discrepancies captured by the SPR and EHR data, respectively. A novel concordance assessment was also developed to understand the detec-tion power and capabilities of SPR and EHR data. Approximately 70% of the SPRs contained valid patient IDs and medication names, making them feasible for data integration. After combining the 2 sources, the investigative team reviewed 2307 medication orders with 10,575 medication administration records (MARs) and 23,397 SPRs. A total of 321 MAR and 682 SPR dis-crepancies were identified, with vasopressors showing the highest discrepancy rates, followed by narcotics and total parenteral nutrition. Compared with EHR MARs, substantial dosing discrepancies were more commonly detectable using the SPRs. The concordance analysis showed little overlap between MAR and SPR discrepan-cies, with most discrepancies captured by the SPR data. We integrated smart infusion pump information with EHR data to analyze the most error-prone phases of the medication lifecycle. The findings suggested that SPRs could be a more reliable data source for medication error detection. Ultimately, it is imperative to integrate SPR information with EHR data to fully detect and mitigate medication administration errors in the clinical setting.
Sections du résumé
BACKGROUND
BACKGROUND
At present, electronic health records (EHRs) are the central focus of clinical informatics given their role as the primary source of clinical data. Despite their granularity, the EHR data heavily rely on manual input and are prone to human errors. Many other sources of data exist in the clinical setting, including digital medical devices such as smart infusion pumps. When incorporated with prescribing data from EHRs, smart pump records (SPRs) are capable of shedding light on actions that take place during the medication use process. However, harmoniz-ing the 2 sources is hindered by multiple technical challenges, and the data quality and utility of SPRs have not been fully realized.
OBJECTIVE
OBJECTIVE
This study aims to evaluate the quality and utility of SPRs incorporated with EHR data in detecting medication administration errors. Our overarching hypothesis is that SPRs would contribute unique information in the med-ication use process, enabling more comprehensive detection of discrepancies and potential errors in medication administration.
METHODS
METHODS
We evaluated the medication use process of 9 high-risk medications for patients admitted to the neonatal inten-sive care unit during a 1-year period. An automated algorithm was developed to align SPRs with their medica-tion orders in the EHRs using patient ID, medication name, and timestamp. The aligned data were manually re-viewed by a clinical research coordinator and 2 pediatric physicians to identify discrepancies in medication ad-ministration. The data quality of SPRs was assessed with the proportion of information that was linked to valid EHR orders. To evaluate their utility, we compared the frequency and severity of discrepancies captured by the SPR and EHR data, respectively. A novel concordance assessment was also developed to understand the detec-tion power and capabilities of SPR and EHR data.
RESULTS
RESULTS
Approximately 70% of the SPRs contained valid patient IDs and medication names, making them feasible for data integration. After combining the 2 sources, the investigative team reviewed 2307 medication orders with 10,575 medication administration records (MARs) and 23,397 SPRs. A total of 321 MAR and 682 SPR dis-crepancies were identified, with vasopressors showing the highest discrepancy rates, followed by narcotics and total parenteral nutrition. Compared with EHR MARs, substantial dosing discrepancies were more commonly detectable using the SPRs. The concordance analysis showed little overlap between MAR and SPR discrepan-cies, with most discrepancies captured by the SPR data.
CONCLUSIONS
CONCLUSIONS
We integrated smart infusion pump information with EHR data to analyze the most error-prone phases of the medication lifecycle. The findings suggested that SPRs could be a more reliable data source for medication error detection. Ultimately, it is imperative to integrate SPR information with EHR data to fully detect and mitigate medication administration errors in the clinical setting.
Identifiants
pubmed: 32876578
pii: v8i9e19774
doi: 10.2196/19774
pmc: PMC7495258
doi:
Types de publication
Journal Article
Langues
eng
Pagination
e19774Subventions
Organisme : NLM NIH HHS
ID : R01 LM012230
Pays : United States
Informations de copyright
©Yizhao Ni, Todd Lingren, Hannah Huth, Kristen Timmons, Krisin Melton, Eric Kirkendall. Originally published in JMIR Medical Informatics (http://medinform.jmir.org), 02.09.2020.
Références
Biochem Med (Zagreb). 2012;22(3):276-82
pubmed: 23092060
BMC Med Inform Decis Mak. 2019 Nov 7;19(1):213
pubmed: 31699078
J Biomed Inform. 2015 Oct;57:124-33
pubmed: 26190267
J Intraven Nurs. 1997 Nov-Dec;20(6):311-6
pubmed: 9423393
J Am Med Inform Assoc. 2014 Sep-Oct;21(5):776-84
pubmed: 24401171
Am J Health Syst Pharm. 2013 Apr 15;70(8):680-6
pubmed: 23552046
Qual Saf Health Care. 2010 Oct;19(5):430-4
pubmed: 20427310
J Am Med Inform Assoc. 2018 May 1;25(5):555-563
pubmed: 29329456
J Med Internet Res. 2019 May 22;21(5):e13047
pubmed: 31120022
Crit Care Med. 2005 Mar;33(3):533-40
pubmed: 15753744
Pharmacotherapy. 2018 Aug;38(8):842-850
pubmed: 29883535
BMJ Qual Saf. 2017 Feb;26(2):131-140
pubmed: 26908900
Jt Comm J Qual Patient Saf. 2012 Mar;38(3):112-9
pubmed: 22435228
BMJ Qual Saf. 2011 Dec;20(12):1027-34
pubmed: 21690248
Am J Health Syst Pharm. 2018 Jul 15;75(14):1064-1068
pubmed: 29987060
BMJ Qual Saf. 2018 Nov;27(11):892-901
pubmed: 29627799
Jt Comm J Qual Patient Saf. 2019 Oct;45(10):649-661
pubmed: 31500950
Pharm World Sci. 2003 Jun;25(3):112-7
pubmed: 12840964
Pediatrics. 2006 Nov;118(5):1872-9
pubmed: 17079557
Drug Saf. 2014 Dec;37(12):1011-20
pubmed: 25294653
J Pharm Pharmacol. 2017 May;69(5):529-536
pubmed: 27334458
Appl Clin Inform. 2014 Jan 08;5(1):25-45
pubmed: 24734122
Qual Saf Health Care. 2005 Apr;14(2):80-6
pubmed: 15805451