Radiation dose management systems-requirements and recommendations for users from the ESR EuroSafe Imaging initiative.
Dose management systems
Optimisation
Quality assurance
Radiation protection
Journal
European radiology
ISSN: 1432-1084
Titre abrégé: Eur Radiol
Pays: Germany
ID NLM: 9114774
Informations de publication
Date de publication:
Apr 2021
Apr 2021
Historique:
received:
21
08
2020
accepted:
11
09
2020
revised:
07
09
2020
pubmed:
23
9
2020
medline:
15
4
2021
entrez:
22
9
2020
Statut:
ppublish
Résumé
The European Directive 2013/59/Euratom requires member states of the European Union to ensure justification and optimisation of radiological procedures and store information on patient exposure for analysis and quality assurance. The EuroSafe Imaging campaign of the European Society of Radiology created a working group (WG) on "Dose Management" with the aim to provide European recommendations on the implementation of dose management systems (DMS) in clinical practice. The WG follows Action 4: "Promote dose management systems to establish local, national, and European diagnostic reference levels (DRL)" of the EuroSafe Imaging Call for Action 2018. DMS are designed for medical practitioners, radiographers, medical physics experts (MPE) and other health professionals involved in imaging to support their tasks and duties of radiation protection in accordance with local and national requirements. The WG analysed requirements and critical points when installing a DMS and classified the individual functions at different performance levels. KEY POINTS: • DMS are very helpful software tools for monitoring patient exposure, optimisation, compliance with DRLs and quality assurance. • DMS can help to fulfil dosimetric aspects of the European Directive 2013/59/Euratom. • The EuroSafe WG analyses DMS requirements and gives recommendations for users.
Identifiants
pubmed: 32959080
doi: 10.1007/s00330-020-07290-x
pii: 10.1007/s00330-020-07290-x
pmc: PMC7979596
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
2106-2114Références
European Council Directive 2013/59/Euratom on basic safety standards for protection against the dangers arising from exposure to ionising radiation (2014). Official Journal of the European Union L13;57:1-73 Available at https://eur-lex.europa.eu/LexUriServ/LexUriServ.do?uri=OJ:L:2014:013:0001:0073:EN:PDF . Accessed 13 Jan 2020
European Society of Radiology (ESR) (2015) Summary of the European Directive 2013/59/Euratom: essentials for health professionals in radiology. Insights Imaging 6:411–417
doi: 10.1007/s13244-015-0410-4
EuroSafe Imaging Call for Action 2018, http://www.eurosafeimaging.org/about/call-for-action Accessed 13 Jan 2020
Vañó E, Miller DL, Martin CJ et al (2017) ICRP Publication 135: Diagnostic reference levels in medical imaging. Ann ICRP 46(1):1–144. https://doi.org/10.1177/0146645317717209
Radiation Protection No. 185 (2018) European guidelines on diagnostic reference levels for paediatric imaging. Eur Commission. https://doi.org/10.2833/486256
European Society of Radiology (ESR) (2013) How to manage accidental and unintended exposure in radiology: an ESR white paper. Insights Imaging 10:23. https://doi.org/10.1186/s13244-019-0691-0
Pyfferoen L, Mulkens TH, Zanca F, De Bondt T, Parizel PM, Casselman JW (2017) Benchmarking adult CT-dose levels to regional and national references using a dose-tracking software: a multicentre experience. Insights Imaging 8:513–521
Ruiz-Cruces R, Vano E, Carrera-Magariño F et al (2016) Diagnostic reference levels and complexity indices in interventional radiology: a national programme. Eur Radiol 26(12):4268–4276. https://doi.org/10.1007/s00330-016-4334-2
doi: 10.1007/s00330-016-4334-2
pubmed: 27384609
ESR iGuide, Clinical Decision Support using European Imaging Referral Guidelines. https://www.myesr.org/esriguide , Accessed 13 Jan 2020
Brambilla M, Vassileva J, Kuchcinska A, Rehani MM (2019) Multinational data on cumulative radiation exposure of patients from recurrent radiological procedures: call for action. Eur Radiol 30. https://doi.org/10.1007/s00330-019-06528-7