Fast In Situ Image Reconstruction for Proton Radiography.

Patient Alignment Proton Computed Tomography Proton Radiography Range Verification Relative Stopping Power Treatment Planning

Journal

Journal of radiation oncology
ISSN: 1948-7894
Titre abrégé: J Radiat Oncol
Pays: Germany
ID NLM: 101590774

Informations de publication

Date de publication:
Jun 2019
Historique:
entrez: 6 11 2020
pubmed: 1 6 2019
medline: 1 6 2019
Statut: ppublish

Résumé

Proton beam therapy is an emerging modality for cancer treatment that, compared to X-ray radiation therapy, promises to provide better dose delivery to clinical targets with lower doses to normal tissues. Crucial to accurate treatment planning and dose delivery is knowledge of the water equivalent path length (WEPL) of each ray, or pencil beam, from the skin to every point in the target. For protons, this length is estimated from relative stopping power based on X-ray Hounsfield units. Unfortunately, such estimates lead to 3 to 4% uncertainties in the proton range prediction. Therefore, protons in the Bragg peak may overshoot (or undershoot) the desired stopping depth in the target causing tissue damage beyond the target volume. Recent studies indicate that tomographic imaging using protons has the potential to provide directly more accurate measurement of RSPs with significantly lower radiation dose than X-rays. We are currently working on a proton radiography system that promises to provide accurate two-dimensional (2D) images of WEPL values for protons that pass through the body. These will be suitable for positioning and range verification in daily treatments. In this study, we demonstrate that this system is capable of rapidly achieving such accurate images in clinically meaningful times. We have developed a software platform to characterize the potential performance of the prototype proton radiography system. We use Geant4 to simulate raw data detected by the device. An especially-written software - pRad - was written to process these data as they are received and uses iterative methods to generate radiographs. The software has been designed to generate a radiograph from a few million protons in under a minute after receiving the first proton from the device. We used a head phantom with known chemical compositions that could be modelled quite accurately in Geant4 simulations of proton radiographs. The radiographs are displayed as pixelated WEPL values displayed on a 2D gray scale image of WEPL values. Rapid radiograph reconstruction of 3D phantoms using simulated proton pencil beams have been achieved with our software platform. On a modest desktop computer with a single central processing unit (CPU) and a single graphics processing unit (GPU), it takes about 11 seconds to reconstruct images using iterative linear algorithms to reconstruct a radiograph from 7.6 million protons. For the radiographic reconstructions of the head phantom described here, the mean WEPL errors, in the proton radiograph using a large majority of the pixels in the complete image were less than 1 mm when compared to images obtained without proton scattering and without detector resolution included. We have demonstrated, through computer simulations of proton irradiation of a pediatric head phantom using the newly built pRad detector and image reconstruction software, that high quality proton radiographs can be generated for patient alignment and verification of water equivalent thickness of the patient before each treatment.

Identifiants

pubmed: 33154789
doi: 10.1007/s13566-019-00387-x
pmc: PMC7641509
mid: NIHMS1530267
doi:

Types de publication

Journal Article

Langues

eng

Pagination

185-198

Subventions

Organisme : NCI NIH HHS
ID : R44 CA203499
Pays : United States

Déclaration de conflit d'intérêts

Conflict of Interest Statement The authors have intellectual property rights to the innovations described in this paper. James S. Welsh has served as a medical advisor to ProTom International. Don F. Dejongh is a co-owner of ProtonVDA Inc.

Références

Phys Med Biol. 2017 Aug 18;62(17):L20-L30
pubmed: 28742053
Phys Med Biol. 2016 Dec 7;61(23):8232-8248
pubmed: 27811399
Phys Med Biol. 2004 Jul 7;49(13):2899-911
pubmed: 15285255
Med Phys. 2005 Apr;32(4):1035-46
pubmed: 15895588
Med Phys. 2017 Jun;44(6):2332-2344
pubmed: 28295434
Phys Med Biol. 2013 Aug 7;58(15):R131-60
pubmed: 23863203
Med Phys. 2017 Mar;44(3):1002-1016
pubmed: 28094862
Acta Oncol. 2015;54(9):1254-8
pubmed: 26329760
Med Phys. 2008 Nov;35(11):4849-56
pubmed: 19070218
J Theor Biol. 1970 Dec;29(3):471-81
pubmed: 5492997
J Radiat Oncol. 2019 Mar;8(1):97-101
pubmed: 33042477
Int J Radiat Oncol Biol Phys. 2017 Sep 1;99(1):210-218
pubmed: 28816148
Phys Med Biol. 2016 Aug 21;61(16):6203-30
pubmed: 27476548

Auteurs

Caesar E Ordoñez (CE)

Northern Illinois University, Center for Research Computing and Data, DeKalb, IL, 60115, USA.

Nicholas T Karonis (NT)

Northern Illinois University, Computer Science Department, DeKalb, IL, 60115, USA.
Argonne National Laboratory, Data Science and Learning Division, Argonne, IL, 60439, USA.

Kirk L Duffin (KL)

Northern Illinois University, Computer Science Department, DeKalb, IL, 60115, USA.

John R Winans (JR)

Northern Illinois University, Center for Research Computing and Data, DeKalb, IL, 60115, USA.

Ethan A DeJongh (EA)

ProtonVDA, Inc., Naperville, IL, 60563, USA.

Don F DeJongh (DF)

ProtonVDA, Inc., Naperville, IL, 60563, USA.

George Coutrakon (G)

Northern Illinois University, Physics Department, DeKalb, IL, 60115, USA.

Nicole F Myers (NF)

Northern Illinois University, Computer Science Department, DeKalb, IL, 60115, USA.

Mark Pankuch (M)

Northwestern Medicine Chicago Proton Center, Warrenville, IL, 60555, USA.

James S Welsh (JS)

Edward Hines Jr VA Medical Center, Radiation Oncology Service, Hines, IL, 60141, USA.
Loyola University Stritch School of Medicine, Department of Radiation Oncology, Maywood, IL 60153, USA.

Classifications MeSH