Use of porous custom-made cones for meta-diaphyseal bone defects reconstruction in knee revision surgery: a clinical and biomechanical analysis.


Journal

Archives of orthopaedic and trauma surgery
ISSN: 1434-3916
Titre abrégé: Arch Orthop Trauma Surg
Pays: Germany
ID NLM: 9011043

Informations de publication

Date de publication:
Dec 2020
Historique:
received: 26 05 2020
accepted: 15 10 2020
pubmed: 11 11 2020
medline: 27 4 2021
entrez: 10 11 2020
Statut: ppublish

Résumé

Although the practice of metaphyseal reconstruction has obtained successful clinical and radiological results in revision total knee surgery, off-the-shelf devices aren't an effective solution for all patients as they do not cover the full range of clinical possibilities. For this reason, during severe knee revisions, custom-made porous titanium cementless metaphyseal cones are nowadays employed as alternative to traditional surgeries. The aim of this study is to understand the benefits gained by the use of the custom-made cones against the performance of more traditional techniques, such as the use of cemented or cementless stems. Thus, a retrospective study on eleven patients and a biomechanical finite element analysis (FEA) was developed, based upon three clinical cases of the clinical analyzed cohort. Eleven patients underwent staged total knee arthroplasty revision with the use of 16 custom-made cones to correct severe femoral and tibial meta-diaphyseal bone defects. Clinical scores and range of movement were observed during the follow-up period (mean follow-up 26 ± 9.4 months). Reason for surgery was periprosthetic joint infection (PJI) in eight patients and post-traumatic osteomyelitis in the other three patients. Three patients previously affected by PJI were selected among the eleven patients of the clinical population. For those patients, bone geometries and implants during surgery were replicated in silico and analyzed during different daily activities. For the same patients, as alternative solution for surgery, the use of cemented or cementless stems was also simulated by FEA. Stress patterns in different region of interest and risk of fracture in the bone were calculated and compared. No loosening, component migration, or mismatches between preoperative planning and intraoperative findings were clinically registered. Biomechanical results demonstrated that the use of custom-made cones induces a more homogeneously distributed bone stress than the other two techniques that concentrate the stress in spotted regions. The risk of fracture is comparable between the use of custom-made cones and cemented technique, while press-fit configurations increase the risk of fracture (more than 35%). Based upon the clinical evidence and the findings after the FEAs, the practice of porous custom-made metaphyseal cones in severe revisions of knee arthroplasties is showing promising biomechanical results. The homogeneous stresses distributions and the lower bone stress gradient could justify a reduction of bone fractures and the risk of implant loosening which could be the explanation to the successful clinical outcomes.

Identifiants

pubmed: 33170352
doi: 10.1007/s00402-020-03670-6
pii: 10.1007/s00402-020-03670-6
doi:

Substances chimiques

Titanium D1JT611TNE

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

2041-2055

Subventions

Organisme : Fonds De La Recherche Scientifique - FNRS
ID : CDR 19545501
Organisme : Fonds De La Recherche Scientifique - FNRS
ID : CDR 29155446
Organisme : Université Libre de Bruxelles
ID : FER 2017

Auteurs

Giorgio Burastero (G)

Orthopedic and Traumatology Unit 2, Prosthetic Surgery, Santa Corona Hospital, Pietra Ligure, Italy.

Silvia Pianigiani (S)

Adler Ortho, Cormano, MI, Italy.

Cristiana Zanvettor (C)

BEAMS Department (Bio Electro and Mechanical Systems), École Polytechnique de Bruxelles, Université Libre de Bruxelles, Av. F. Roosevelt, 50 CP165/56, 1050, Brussels, Belgium.

Luca Cavagnaro (L)

Orthopedic and Traumatology Unit 2, Prosthetic Surgery, Santa Corona Hospital, Pietra Ligure, Italy.

Francesco Chiarlone (F)

Orthopedic and Traumatology Unit 2, Prosthetic Surgery, Santa Corona Hospital, Pietra Ligure, Italy.

Bernardo Innocenti (B)

BEAMS Department (Bio Electro and Mechanical Systems), École Polytechnique de Bruxelles, Université Libre de Bruxelles, Av. F. Roosevelt, 50 CP165/56, 1050, Brussels, Belgium. bernardo.innocenti@ulb.be.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH