Evaluation of X-ray table mattresses for radiation attenuation and impact on image quality.
Attenuation
Image quality
Mattress
Radiography
Journal
Radiography (London, England : 1995)
ISSN: 1532-2831
Titre abrégé: Radiography (Lond)
Pays: Netherlands
ID NLM: 9604102
Informations de publication
Date de publication:
02 2021
02 2021
Historique:
received:
10
07
2020
revised:
23
10
2020
accepted:
24
10
2020
pubmed:
14
11
2020
medline:
30
9
2021
entrez:
13
11
2020
Statut:
ppublish
Résumé
Mattresses in the radiology department tend to be an overlooked aspect of imaging equipment. This paper evaluates the radiation attenuation characteristics of mattresses and the effect they have on image quality. Thirteen mattresses (from new to 20 years of age) were evaluated. Incident air kerma (IAK) was measured in two conditions, with and without mattress over a range of exposure factors using a digital dosimeter. Image quality was assessed by calculating the inverse image quality factor (IQF Measured IAK and image quality was affected by the addition of a mattress with older mattresses having greater attenuation; there is a moderate/large correlation (0.38-0.51) between age and IAK. IQF The results indicate that while the presence of a mattress does impact on transmitted radiation and the quality of the image, the clinical impact is insignificant. Attenuation correlates with age but with no clinical significance. There is no correlation between age and image quality. Quality control tests for attenuation and impact on image quality are not required in clinical practice. The method could be used by manufacturers to test new materials and mattresses and could provide users with specifications of new products.
Identifiants
pubmed: 33183977
pii: S1078-8174(20)30224-8
doi: 10.1016/j.radi.2020.10.014
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
215-220Informations de copyright
Crown Copyright © 2020. Published by Elsevier Ltd. All rights reserved.