Fedora-type magnetic compression anastomosis device for intestinal anastomosis.

Anastomotic stenosis Compression pressure Fedora-type magnetic compression anastomosis device Magnetic compression anastomosis Size of anastomat

Journal

World journal of gastroenterology
ISSN: 2219-2840
Titre abrégé: World J Gastroenterol
Pays: United States
ID NLM: 100883448

Informations de publication

Date de publication:
14 Nov 2020
Historique:
received: 12 08 2020
revised: 29 08 2020
accepted: 10 09 2020
entrez: 3 12 2020
pubmed: 4 12 2020
medline: 15 5 2021
Statut: ppublish

Résumé

Although previous studies have confirmed the feasibility of magnetic compression anastomosis (MCA), there is still a risk of long-term anastomotic stenosis. For traditional MCA devices, a large device is associated with great pressure, and eventually increased leakage. To develop a novel MCA device to simultaneously meet the requirements of pressure and size. Traditional nummular MCA devices of all possible sizes were used to conduct ileac anastomosis in rats. The mean (± SD) circumference of the ileum was 13.34 ± 0.12 mm. Based on short- and long-term follow-up results, we determined the appropriate pressure range and minimum size. Thereafter, we introduced a novel "fedora-type" MCA device, which entailed the use of a nummular magnet with a larger sheet metal. With traditional MCA devices, the anastomoses experienced stenosis and even closure during the long-term follow-up when the anastomat was smaller than Φ5 mm. However, the risk of leakage increased when it was larger than Φ4 mm. On comparison of the different designs, it was found that the "fedora-type" MCA device should be composed of a Φ4-mm nummular magnet with a Φ6-mm sheet metal. The diameter of the MCA device should be greater than 120% of the enteric diameter. The novel "fedora-type" MCA device controls the pressure and optimizes the size.

Sections du résumé

BACKGROUND BACKGROUND
Although previous studies have confirmed the feasibility of magnetic compression anastomosis (MCA), there is still a risk of long-term anastomotic stenosis. For traditional MCA devices, a large device is associated with great pressure, and eventually increased leakage.
AIM OBJECTIVE
To develop a novel MCA device to simultaneously meet the requirements of pressure and size.
METHODS METHODS
Traditional nummular MCA devices of all possible sizes were used to conduct ileac anastomosis in rats. The mean (± SD) circumference of the ileum was 13.34 ± 0.12 mm. Based on short- and long-term follow-up results, we determined the appropriate pressure range and minimum size. Thereafter, we introduced a novel "fedora-type" MCA device, which entailed the use of a nummular magnet with a larger sheet metal.
RESULTS RESULTS
With traditional MCA devices, the anastomoses experienced stenosis and even closure during the long-term follow-up when the anastomat was smaller than Φ5 mm. However, the risk of leakage increased when it was larger than Φ4 mm. On comparison of the different designs, it was found that the "fedora-type" MCA device should be composed of a Φ4-mm nummular magnet with a Φ6-mm sheet metal.
CONCLUSION CONCLUSIONS
The diameter of the MCA device should be greater than 120% of the enteric diameter. The novel "fedora-type" MCA device controls the pressure and optimizes the size.

Identifiants

pubmed: 33268950
doi: 10.3748/wjg.v26.i42.6614
pmc: PMC7673965
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

6614-6625

Informations de copyright

©The Author(s) 2020. Published by Baishideng Publishing Group Inc. All rights reserved.

Déclaration de conflit d'intérêts

Conflict-of-interest statement: All authors declare no conflicts of interest related to this article.

Références

J Surg Res. 2018 Nov;231:24-29
pubmed: 30278935
J Pediatr Surg. 2020 Mar;55(3):425-432
pubmed: 31128845
Surg Case Rep. 2018 Aug 15;4(1):95
pubmed: 30112678
World J Surg. 2018 Dec;42(12):4039-4045
pubmed: 29947988
Sci Rep. 2018 Aug 2;8(1):11620
pubmed: 30072707
Surg Laparosc Endosc Percutan Tech. 2019 Oct;29(5):e84-e87
pubmed: 31107851
IEEE Trans Biomed Eng. 2018 Oct;65(10):2178-2189
pubmed: 29989954
Pediatr Int. 2017 Jun;59(6):737-739
pubmed: 28626977
Surg Endosc. 2019 Jun;33(6):1721-1730
pubmed: 30805789
European J Pediatr Surg Rep. 2018 Jan;6(1):e37-e39
pubmed: 29796381
Hepatobiliary Surg Nutr. 2019 Apr;8(2):177-178
pubmed: 31098373
Surg Case Rep. 2017 Dec;3(1):73
pubmed: 28547740
Gastrointest Endosc. 2017 Nov;86(5):904-912
pubmed: 28716404
Surg Innov. 2014 Feb;21(1):65-73
pubmed: 23592733
Surg Innov. 2017 Oct;24(5):432-439
pubmed: 28745145
J Surg Res. 2020 Jan;245:249-256
pubmed: 31421370
Clin Otolaryngol. 2019 Nov;44(6):1214-1217
pubmed: 30968566
J Gastroenterol Hepatol. 2019 Mar;34(3):589-594
pubmed: 30278106
J Laparoendosc Adv Surg Tech A. 2017 Dec;27(12):1314-1317
pubmed: 28976806
J Pediatr Surg. 2019 Mar;54(3):429-433
pubmed: 30309731
Surg Endosc. 2020 Jun;34(6):2541-2550
pubmed: 31399950
Surg Endosc. 2018 Feb;32(2):779-789
pubmed: 28779259
World J Gastroenterol. 2019 Aug 14;25(30):4213-4221
pubmed: 31435174
J Pediatr Surg. 2019 Oct;54(10):2166-2171
pubmed: 30929946
Dis Colon Rectum. 1991 Jun;34(6):442-8
pubmed: 1953849
World J Gastroenterol. 2016 Feb 21;22(7):2326-35
pubmed: 26900294
J Surg Res. 2019 Jul;239:52-59
pubmed: 30802705
J Laparoendosc Adv Surg Tech A. 2019 Jun;29(6):852-855
pubmed: 30882275
Neurol Med Chir (Tokyo). 1980 May;20(5):497-505
pubmed: 6157124
Gastroenterology Res. 2017 Aug;10(4):255-258
pubmed: 28912914

Auteurs

Huan Chen (H)

National Local Joint Engineering Research Center for Precision Surgery & Regenerative Medicine, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710061, Shaanxi Province, China.

Tao Ma (T)

National Local Joint Engineering Research Center for Precision Surgery & Regenerative Medicine, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710061, Shaanxi Province, China.

Yue Wang (Y)

National Local Joint Engineering Research Center for Precision Surgery & Regenerative Medicine, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710061, Shaanxi Province, China.

Hao-Yang Zhu (HY)

National Local Joint Engineering Research Center for Precision Surgery & Regenerative Medicine, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710061, Shaanxi Province, China.

Zhe Feng (Z)

National Local Joint Engineering Research Center for Precision Surgery & Regenerative Medicine, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710061, Shaanxi Province, China.

Rong-Qian Wu (RQ)

National Local Joint Engineering Research Center for Precision Surgery & Regenerative Medicine, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710061, Shaanxi Province, China.

Yi Lv (Y)

National Local Joint Engineering Research Center for Precision Surgery & Regenerative Medicine, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710061, Shaanxi Province, China.

Ding-Hui Dong (DH)

National Local Joint Engineering Research Center for Precision Surgery & Regenerative Medicine, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710061, Shaanxi Province, China.

Articles similaires

Robotic Surgical Procedures Animals Humans Telemedicine Models, Animal

Odour generalisation and detection dog training.

Lyn Caldicott, Thomas W Pike, Helen E Zulch et al.
1.00
Animals Odorants Dogs Generalization, Psychological Smell
Animals TOR Serine-Threonine Kinases Colorectal Neoplasms Colitis Mice
Animals Tail Swine Behavior, Animal Animal Husbandry

Classifications MeSH