Mortality with Paclitaxel-Coated Devices in Peripheral Artery Disease.
Journal
The New England journal of medicine
ISSN: 1533-4406
Titre abrégé: N Engl J Med
Pays: United States
ID NLM: 0255562
Informations de publication
Date de publication:
24 12 2020
24 12 2020
Historique:
pubmed:
10
12
2020
medline:
15
1
2021
entrez:
9
12
2020
Statut:
ppublish
Résumé
The results of a recent meta-analysis aroused concern about an increased risk of death associated with the use of paclitaxel-coated angioplasty balloons and stents in lower-limb endovascular interventions for symptomatic peripheral artery disease. We conducted an unplanned interim analysis of data from a multicenter, randomized, open-label, registry-based clinical trial. At the time of the analysis, 2289 patients had been randomly assigned to treatment with drug-coated devices (the drug-coated-device group, 1149 patients) or treatment with uncoated devices (the uncoated-device group, 1140 patients). Randomization was stratified according to disease severity on the basis of whether patients had chronic limb-threatening ischemia (1480 patients) or intermittent claudication (809 patients). The single end point for this interim analysis was all-cause mortality. No patients were lost to follow-up. Paclitaxel was used as the coating agent for all the drug-coated devices. During a mean follow-up of 2.49 years, 574 patients died, including 293 patients (25.5%) in the drug-coated-device group and 281 patients (24.6%) in the uncoated-device group (hazard ratio, 1.06; 95% confidence interval, 0.92 to 1.22). At 1 year, all-cause mortality was 10.2% (117 patients) in the drug-coated-device group and 9.9% (113 patients) in the uncoated-device group. During the entire follow-up period, there was no significant difference in the incidence of death between the treatment groups among patients with chronic limb-threatening ischemia (33.4% [249 patients] in the drug-coated-device group and 33.1% [243 patients] in the uncoated-device group) or among those with intermittent claudication (10.9% [44 patients] and 9.4% [38 patients], respectively). In this randomized trial in which patients with peripheral artery disease received treatment with paclitaxel-coated or uncoated endovascular devices, the results of an unplanned interim analysis of all-cause mortality did not show a difference between the groups in the incidence of death during 1 to 4 years of follow-up. (Funded by the Swedish Research Council and others; ClinicalTrials.gov number, NCT02051088.).
Sections du résumé
BACKGROUND
The results of a recent meta-analysis aroused concern about an increased risk of death associated with the use of paclitaxel-coated angioplasty balloons and stents in lower-limb endovascular interventions for symptomatic peripheral artery disease.
METHODS
We conducted an unplanned interim analysis of data from a multicenter, randomized, open-label, registry-based clinical trial. At the time of the analysis, 2289 patients had been randomly assigned to treatment with drug-coated devices (the drug-coated-device group, 1149 patients) or treatment with uncoated devices (the uncoated-device group, 1140 patients). Randomization was stratified according to disease severity on the basis of whether patients had chronic limb-threatening ischemia (1480 patients) or intermittent claudication (809 patients). The single end point for this interim analysis was all-cause mortality.
RESULTS
No patients were lost to follow-up. Paclitaxel was used as the coating agent for all the drug-coated devices. During a mean follow-up of 2.49 years, 574 patients died, including 293 patients (25.5%) in the drug-coated-device group and 281 patients (24.6%) in the uncoated-device group (hazard ratio, 1.06; 95% confidence interval, 0.92 to 1.22). At 1 year, all-cause mortality was 10.2% (117 patients) in the drug-coated-device group and 9.9% (113 patients) in the uncoated-device group. During the entire follow-up period, there was no significant difference in the incidence of death between the treatment groups among patients with chronic limb-threatening ischemia (33.4% [249 patients] in the drug-coated-device group and 33.1% [243 patients] in the uncoated-device group) or among those with intermittent claudication (10.9% [44 patients] and 9.4% [38 patients], respectively).
CONCLUSIONS
In this randomized trial in which patients with peripheral artery disease received treatment with paclitaxel-coated or uncoated endovascular devices, the results of an unplanned interim analysis of all-cause mortality did not show a difference between the groups in the incidence of death during 1 to 4 years of follow-up. (Funded by the Swedish Research Council and others; ClinicalTrials.gov number, NCT02051088.).
Identifiants
pubmed: 33296560
doi: 10.1056/NEJMoa2005206
doi:
Substances chimiques
Paclitaxel
P88XT4IS4D
Banques de données
ClinicalTrials.gov
['NCT02051088']
Types de publication
Comparative Study
Journal Article
Multicenter Study
Randomized Controlled Trial
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM
Pagination
2538-2546Subventions
Organisme : Hjärt-Lungfonden
ID : 20130446
Pays : International
Organisme : Västra Götalandsregionen
ID : ALFGBG 77190 and Regional 205531; ALFGBG 822921 an
Pays : International
Organisme : Vetenskapsrådet
ID : 521-2013-3436 and 921-2014-7231
Pays : International
Informations de copyright
Copyright © 2020 Massachusetts Medical Society.