Enhanced detection and sizing algorithm to improve LOAC optical particle counter performances.
Journal
Applied optics
ISSN: 1539-4522
Titre abrégé: Appl Opt
Pays: United States
ID NLM: 0247660
Informations de publication
Date de publication:
01 Dec 2020
01 Dec 2020
Historique:
entrez:
28
12
2020
pubmed:
29
12
2020
medline:
29
12
2020
Statut:
ppublish
Résumé
The light optical aerosols counter (LOAC) instrument is an optical aerosol counter that allows atmospheric particles from 0.2 to 50 µm to be individually counted and classified by size. The scattered light due to the interaction between a particle and a laser beam is captured by photodiodes and produces an electric pulse. Detecting these pulses and measuring their amplitude is a key process to measure the particle size. This paper first details the current method to achieve such estimation and proposes a correlation algorithm to improve the detection and the assessment of particle sizes. In the last section, some test results are presented to demonstrate the performance of this new algorithm.
Identifiants
pubmed: 33361910
pii: 444052
doi: 10.1364/AO.408959
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM