Broadband high-energy resolution hard x-ray spectroscopy using transition edge sensors at SPring-8.
Journal
The Review of scientific instruments
ISSN: 1089-7623
Titre abrégé: Rev Sci Instrum
Pays: United States
ID NLM: 0405571
Informations de publication
Date de publication:
01 Jan 2021
01 Jan 2021
Historique:
entrez:
30
1
2021
pubmed:
31
1
2021
medline:
31
1
2021
Statut:
ppublish
Résumé
We have succeeded in operating a transition-edge sensor (TES) spectrometer and evaluating its performance at the SPring-8 synchrotron x-ray light source. The TES spectrometer consists of a 240 pixel National Institute of Standards and Technology (NIST) TES system, and 220 pixels are operated simultaneously with an energy resolution of 4 eV at 6 keV at a rate of ∼1 c/s pixel