Quantitative Assessment of Weight-Bearing Fracture Biomechanics Using Extremity Cone-Beam CT.

3D-2D registration deformable registration fracture biomechanics orthopedic imaging quantitative imaging surgical hardware

Journal

Proceedings of SPIE--the International Society for Optical Engineering
ISSN: 0277-786X
Titre abrégé: Proc SPIE Int Soc Opt Eng
Pays: United States
ID NLM: 101524122

Informations de publication

Date de publication:
Feb 2020
Historique:
entrez: 22 2 2021
pubmed: 23 2 2021
medline: 23 2 2021
Statut: ppublish

Résumé

We investigate an application of multisource extremity Cone-Beam CT (CBCT) with capability of weight-bearing tomographic imaging to obtain quantitative measurements of load-induced deformation of metal internal fixation hardware (e.g. tibial plate). Such measurements are desirable to improve the detection of delayed fusion or non-union of fractures, potentially facilitating earlier return to weight-bearing activities. To measure the deformation, we perform a deformable 3D-2D registration of a prior model of the implant to its CBCT projections under load-bearing. This Known-Component Registration (KC-Reg) framework avoids potential errors that emerge when the deformation is estimated directly from 3D reconstructions with metal artifacts. The 3D-2D registration involves a free-form deformable (FFD) point cloud model of the implant and a 3D cubic B-spline representation of the deformation. Gradient correlation is used as the optimization metric for the registration. The proposed approach was tested in experimental studies on the extremity CBCT system. A custom jig was designed to apply controlled axial loads to a fracture model, emulating weight-bearing imaging scenarios. Performance evaluation involved a Sawbone tibia phantom with an ~4 mm fracture gap. The model was fixed with a locking plate and imaged under five loading conditions. To investigate performance in the presence of confounding background gradients, additional experiments were performed with a pre-deformed femoral plate placed in a water bath with Ca bone mineral density inserts. Errors were measured using eight reference BBs for the tibial plate, and surface point distances for the femoral plate, where a prior model of deformed implant was available for comparison. Both in the loaded tibial plate case and for the femoral plate with confounding background gradients, the proposed KC-Reg framework estimated implant deformations with errors of <0.2 mm for the majority of the investigated deformation magnitudes (error range 0.14 - 0.25 mm). The accuracy was comparable between 3D-2D registrations performed from 12 x-ray views and registrations obtained from as few as 3 views. This was likely enabled by the unique three-source x-ray unit on the extremity CBCT scanner, which implements two off-central-plane focal spots that provided oblique views of the field-of-view to aid implant pose estimation. Accurate measurements of fracture hardware deformations under physiological weight-bearing are feasible using an extremity CBCT scanner and FFD 3D-2D registration. The resulting deformed implant models can be incorporated into tomographic reconstructions to reduce metal artifacts and improve quantification of the mineral content of fracture callus in CBCT volumes.

Identifiants

pubmed: 33612913
doi: 10.1117/12.2549768
pmc: PMC7891844
mid: NIHMS1666930
pii:
doi:

Types de publication

Journal Article

Langues

eng

Subventions

Organisme : NIBIB NIH HHS
ID : R01 EB025470
Pays : United States

Références

Proc SPIE Int Soc Opt Eng. 2019 Jun;11072:
pubmed: 31359904
IEEE Trans Med Imaging. 1999 Aug;18(8):712-21
pubmed: 10534053
IEEE Trans Med Imaging. 2012 Oct;31(10):1837-48
pubmed: 22614574
Proc SPIE Int Soc Opt Eng. 2013;8668:86681L
pubmed: 24949189
J Image Guid Surg. 1995;1(2):88-93
pubmed: 9079432
IEEE Trans Pattern Anal Mach Intell. 2010 Dec;32(12):2262-75
pubmed: 20975122
Phys Med Biol. 2017 Apr 21;62(8):3330-3351
pubmed: 28233760
Med Phys. 2011 Aug;38(8):4700-13
pubmed: 21928644
Phys Med Biol. 2015 Oct 21;60(20):8007-24
pubmed: 26421941
Med Phys. 2018 Jan;45(1):144-155
pubmed: 29121409
Phys Med Biol. 2016 Jan 7;61(1):296-319
pubmed: 26649783
Neurosurgery. 2001 Apr;48(4):810-6; discussion 816-7
pubmed: 11322441
Phys Med Biol. 2015 Feb 21;60(4):1415-39
pubmed: 25611041

Auteurs

S Z Liu (SZ)

Department of Biomedical Engineering, Johns Hopkins University, Baltimore, MD 21205.

Q Cao (Q)

Department of Biomedical Engineering, Johns Hopkins University, Baltimore, MD 21205.

G M Osgood (GM)

Department of Orthopedic Surgery, Johns Hopkins Hospital, Baltimore, MD 21205.

J H Siewerdsen (JH)

Department of Biomedical Engineering, Johns Hopkins University, Baltimore, MD 21205.
Russell H. Morgan Department of Radiology, Johns Hopkins Hospital, Baltimore, MD 21205.

J W Stayman (JW)

Department of Biomedical Engineering, Johns Hopkins University, Baltimore, MD 21205.

W Zbijewski (W)

Department of Biomedical Engineering, Johns Hopkins University, Baltimore, MD 21205.

Classifications MeSH