The 1-Megapixel pnCCD detector for the Small Quantum Systems Instrument at the European XFEL: system and operation aspects.
X-ray CCD
X-ray detector
free-electron laser
photon detection
pnCCD
Journal
Journal of synchrotron radiation
ISSN: 1600-5775
Titre abrégé: J Synchrotron Radiat
Pays: United States
ID NLM: 9888878
Informations de publication
Date de publication:
01 Mar 2021
01 Mar 2021
Historique:
received:
01
08
2020
accepted:
29
11
2020
entrez:
2
3
2021
pubmed:
3
3
2021
medline:
3
3
2021
Statut:
ppublish
Résumé
The X-ray free-electron lasers that became available during the last decade, like the European XFEL (EuXFEL), place high demands on their instrumentation. Especially at low photon energies below 1 keV, detectors with high sensitivity, and consequently low noise and high quantum efficiency, are required to enable facility users to fully exploit the scientific potential of the photon source. A 1-Megapixel pnCCD detector with a 1024 × 1024 pixel format has been installed and commissioned for imaging applications at the Nano-Sized Quantum System (NQS) station of the Small Quantum System (SQS) instrument at EuXFEL. The instrument is currently operating in the energy range between 0.5 and 3 keV and the NQS station is designed for investigations of the interaction of intense FEL pulses with clusters, nano-particles and small bio-molecules, by combining photo-ion and photo-electron spectroscopy with coherent diffraction imaging techniques. The core of the imaging detector is a pn-type charge coupled device (pnCCD) with a pixel pitch of 75 µm × 75 µm. Depending on the experimental scenario, the pnCCD enables imaging of single photons thanks to its very low electronic noise of 3 e
Identifiants
pubmed: 33650570
pii: S1600577520015659
doi: 10.1107/S1600577520015659
pmc: PMC7941295
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
576-587Subventions
Organisme : Bundesministerium für Bildung, Wissenschaft, Forschung und Technologie
ID : SAW/2017/MBI4
Informations de copyright
open access.
Références
Rev Sci Instrum. 2014 Mar;85(3):033110
pubmed: 24689567
IUCrJ. 2015 Apr 10;2(Pt 3):371-83
pubmed: 25995846
J Synchrotron Radiat. 2019 Jan 1;26(Pt 1):74-82
pubmed: 30655470
J Synchrotron Radiat. 2019 Sep 1;26(Pt 5):1448-1461
pubmed: 31490132