Comparison of Viscoelastic Testing by Rotational Torsion and Harmonic Resonance Methods.
Hemostasis
Instrument validation
TEG
Transfusion
Viscoelastic testing
Journal
American journal of clinical pathology
ISSN: 1943-7722
Titre abrégé: Am J Clin Pathol
Pays: England
ID NLM: 0370470
Informations de publication
Date de publication:
13 Oct 2021
13 Oct 2021
Historique:
pubmed:
13
5
2021
medline:
10
11
2021
entrez:
12
5
2021
Statut:
ppublish
Résumé
To compare the performance of the TEG 5000 and TEG 6S Global Hemostasis cartridge. We reviewed validation data of the TEG 5000 and TEG 6S Global Hemostasis cartridge. The specimens were analyzed in parallel according to the manufacturer's operating instructions. Fifty-four healthy donors and 13 donors with known hemostatic abnormalities were included. The correlations between instrument types were only moderate-the Spearman rank correlations were 0.55, 0.62, 0.64, and 0.72, respectively, for CK R, K, angle, and maximum amplitude (MA) parameters. Using the manufacturer's device-specific reference ranges to classify results as normal/abnormal, there was weak agreement in the qualitative interpretation of all parameters (Cohen's κ for agreement for CK R, K, angle, and MA was 0.418, 0.154, -0.083, and 0.127, respectively). This could lead to discordant transfusion decisions. These findings indicate that the TEG 5000 and TEG 6S may not be used interchangeably.
Identifiants
pubmed: 33978149
pii: 6274449
doi: 10.1093/ajcp/aqab028
doi:
Types de publication
Comparative Study
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
818-828Informations de copyright
© American Society for Clinical Pathology, 2021. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.