Termination of Repeat Testing in Chemical Laboratories Based on Practice Guidelines: Examining the Effect of Rule-Based Repeat Testing in a Transplantation Center.


Journal

Journal of analytical methods in chemistry
ISSN: 2090-8865
Titre abrégé: J Anal Methods Chem
Pays: Egypt
ID NLM: 101575377

Informations de publication

Date de publication:
2021
Historique:
received: 03 03 2021
accepted: 06 05 2021
entrez: 31 5 2021
pubmed: 1 6 2021
medline: 1 6 2021
Statut: epublish

Résumé

Although the automation of instruments has reduced the variability of results and errors of analysis, in some laboratories, repeating a test to confirm its accuracy is still performed for critical and noncritical results. However, the importance of repeat testing is not well established yet, and there are no clear criteria for repeating a test. In this cross-sectional study, all repeated tests for 26 biochemical analytes (i.e., albumin, alkaline phosphatase (ALP), alanine aminotransferase (ALT), amylase, aspartate aminotransferase (AST), bilirubin total (BT), bilirubin direct (BD), blood urea nitrogen (BUN), calcium, chloride (Cl), cholesterol total (CholT), creatine kinase (CK), creatinine (Cr), glucose, gamma-glutamyl transferase (GGT), high-density lipoprotein-cholesterol (HDL-c), iron, lactate dehydrogenase (LDH), LDL-c, lipase, magnesium (Mg), phosphorus (Ph), protein total (ProtT), total iron binding capacity (TIBC), triglyceride (TG), and uric acid) were assessed in both critical and noncritical ranges over two consecutive months (routine subjective test repeats in the first month and rule-based repeats in the second month). To determine the usefulness of test repeats, differences between the initial and verified results were compared with the allowable bias, and repeat testing was considered necessary if it exceeded the allowable bias range. All causes of repeat testing, including linearity flags, delta checks, clinically significant values, and critical values, were also documented. All data, including the cause of repeats, initial and verified results, time, and costs in the two consecutive months, were transferred to Microsoft Excel for analysis. For comparison of data between the months, Student's A total of 7714 repeat tests were performed over two consecutive months. Although a significant decline (38%) was found in repeated tests in the second month ( Although repeat testing has been shown to be unnecessary in most cases, having a strict policy for repeat testing appears to be more valuable than avoiding it completely. Each laboratory is advised to establish its own protocol for repeat testing based on its own practice.

Sections du résumé

BACKGROUND BACKGROUND
Although the automation of instruments has reduced the variability of results and errors of analysis, in some laboratories, repeating a test to confirm its accuracy is still performed for critical and noncritical results. However, the importance of repeat testing is not well established yet, and there are no clear criteria for repeating a test.
MATERIALS AND METHODS METHODS
In this cross-sectional study, all repeated tests for 26 biochemical analytes (i.e., albumin, alkaline phosphatase (ALP), alanine aminotransferase (ALT), amylase, aspartate aminotransferase (AST), bilirubin total (BT), bilirubin direct (BD), blood urea nitrogen (BUN), calcium, chloride (Cl), cholesterol total (CholT), creatine kinase (CK), creatinine (Cr), glucose, gamma-glutamyl transferase (GGT), high-density lipoprotein-cholesterol (HDL-c), iron, lactate dehydrogenase (LDH), LDL-c, lipase, magnesium (Mg), phosphorus (Ph), protein total (ProtT), total iron binding capacity (TIBC), triglyceride (TG), and uric acid) were assessed in both critical and noncritical ranges over two consecutive months (routine subjective test repeats in the first month and rule-based repeats in the second month). To determine the usefulness of test repeats, differences between the initial and verified results were compared with the allowable bias, and repeat testing was considered necessary if it exceeded the allowable bias range. All causes of repeat testing, including linearity flags, delta checks, clinically significant values, and critical values, were also documented. All data, including the cause of repeats, initial and verified results, time, and costs in the two consecutive months, were transferred to Microsoft Excel for analysis. For comparison of data between the months, Student's
RESULTS RESULTS
A total of 7714 repeat tests were performed over two consecutive months. Although a significant decline (38%) was found in repeated tests in the second month (
CONCLUSION CONCLUSIONS
Although repeat testing has been shown to be unnecessary in most cases, having a strict policy for repeat testing appears to be more valuable than avoiding it completely. Each laboratory is advised to establish its own protocol for repeat testing based on its own practice.

Identifiants

pubmed: 34055449
doi: 10.1155/2021/9955990
pmc: PMC8137285
doi:

Types de publication

Journal Article

Langues

eng

Pagination

9955990

Informations de copyright

Copyright © 2021 Neda Soleimani et al.

Déclaration de conflit d'intérêts

The authors declare that they have no conflicts of interest.

Références

Iran J Pathol. 2018 Spring;13(2):173-178
pubmed: 30697287
PLoS One. 2013 Nov 19;8(11):e80663
pubmed: 24260448
Am J Clin Pathol. 2012 Jan;137(1):20-5
pubmed: 22180474
Clin Chem. 2002 May;48(5):691-8
pubmed: 11978595
Ann Lab Med. 2012 Jan;32(1):5-16
pubmed: 22259773
J Anal Methods Chem. 2020 Feb 12;2020:9857636
pubmed: 32104611
Arch Pathol Lab Med. 2017 Jun;141(6):813-823
pubmed: 28402169
Clin Chem Lab Med. 2010;48(1):63-6
pubmed: 20047530
Clin Chem Lab Med. 2011 Sep 30;50(5):807-12
pubmed: 21958344
Ann Med Health Sci Res. 2015 Jan-Feb;5(1):8-12
pubmed: 25745569
Clin Chem Lab Med. 2014 Dec;52(12):1739-45
pubmed: 24940715
Crit Rev Clin Lab Sci. 2018 Sep;55(6):420-431
pubmed: 30047297
Ethiop J Health Sci. 2018 Mar;28(2):235-244
pubmed: 29983521
JAMA. 1990 Feb 2;263(5):704-7
pubmed: 2296127
Arch Pathol Lab Med. 2011 Apr;135(4):440-4
pubmed: 21466359
Clin Chem Lab Med. 2009;47(2):143-53
pubmed: 19099525

Auteurs

Neda Soleimani (N)

Department of Pathology, Shiraz Medical School, Shiraz University of Medical Sciences, Shiraz, Iran.
Department of Pathology, Shiraz Transplant Center, Abu Ali Sina Hospital, Shiraz University of Medical Sciences, Shiraz, Iran.

Amir Azadi (A)

Department of Pathology, Shiraz Medical School, Shiraz University of Medical Sciences, Shiraz, Iran.

Mohammad Javad Esmaeili (MJ)

Department of Pathology, Shiraz Transplant Center, Abu Ali Sina Hospital, Shiraz University of Medical Sciences, Shiraz, Iran.

Fatemeh Ghodsi (F)

Department of Pathology, Shiraz Transplant Center, Abu Ali Sina Hospital, Shiraz University of Medical Sciences, Shiraz, Iran.

Reza Ghahramani (R)

Department of Pathology, Shiraz Transplant Center, Abu Ali Sina Hospital, Shiraz University of Medical Sciences, Shiraz, Iran.

Azadeh Hafezi (A)

Department of Pathology, Shiraz Transplant Center, Abu Ali Sina Hospital, Shiraz University of Medical Sciences, Shiraz, Iran.

Tayebeh Hosseyni (T)

Department of Pathology, Shiraz Transplant Center, Abu Ali Sina Hospital, Shiraz University of Medical Sciences, Shiraz, Iran.

Arezoo Arabzadeh (A)

Department of Pathology, Shiraz Transplant Center, Abu Ali Sina Hospital, Shiraz University of Medical Sciences, Shiraz, Iran.

Samira Khajeh (S)

Department of Pathology, Shiraz Transplant Center, Abu Ali Sina Hospital, Shiraz University of Medical Sciences, Shiraz, Iran.

Mahsa Farhadi (M)

Department of Pathology, Shiraz Transplant Center, Abu Ali Sina Hospital, Shiraz University of Medical Sciences, Shiraz, Iran.

Sahand Mohammadzadeh (S)

Department of Pathology, Shiraz Medical School, Shiraz University of Medical Sciences, Shiraz, Iran.
Department of Pathology, Shiraz Transplant Center, Abu Ali Sina Hospital, Shiraz University of Medical Sciences, Shiraz, Iran.

Classifications MeSH