Hip implants can restore anatomical and medialized rotation centres in most cases : a 3D templating study comparing four implantation strategies.

Biomechanics Hip arthroplasty Leg length Offset Templating

Journal

Bone & joint open
ISSN: 2633-1462
Titre abrégé: Bone Jt Open
Pays: England
ID NLM: 101770336

Informations de publication

Date de publication:
Jul 2021
Historique:
entrez: 8 7 2021
pubmed: 9 7 2021
medline: 9 7 2021
Statut: ppublish

Résumé

Hip arthroplasty does not always restore normal anatomy. This is due to inaccurate surgery or lack of stem sizes. We evaluated the aptitude of four total hip arthroplasty systems to restore an anatomical and medialized hip rotation centre. Using 3D templating software in 49 CT scans of non-deformed femora, we virtually implanted: 1) small uncemented calcar-guided stems with two offset options (Optimys, Mathys), 2) uncemented straight stems with two offset options (Summit, DePuy Synthes), 3) cemented undersized stems (Exeter philosophy) with three offset options (CPT, ZimmerBiomet), and 4) cemented line-to-line stems (Kerboul philosophy) with proportional offsets (Centris, Mathys). We measured the distance between the templated and the anatomical and 5 mm medialized hip rotation centre. Both rotation centres could be restored within 5 mm in 94% and 92% of cases, respectively. The cemented undersized stem performed best, combining freedom of stem positioning and a large offset range. The uncemented straight stem performed well because of its large and well-chosen offset range, and despite the need for cortical bone contact limiting stem positioning. The cemented line-to-line stem performed less well due to a small range of sizes and offsets. The uncemented calcar-guided stem performed worst, despite 24 sizes and a large and well-chosen offset range. This was attributed to the calcar curvature restricting the stem insertion depth along the femoral axis. In the majority of non-deformed femora, leg length, offset, and anteversion can be restored accurately with non-modular stems during 3D templating. Failure to restore hip biomechanics is mostly due to surgical inaccuracy. Small calcar guided stems offer no advantage to restore hip biomechanics compared to more traditional designs. Cite this article:

Identifiants

pubmed: 34236237
doi: 10.1302/2633-1462.27.BJO-2021-0065.R1
pmc: PMC8325978
doi:

Types de publication

Journal Article

Langues

eng

Pagination

476-485

Références

J Orthop. 2020 Mar 28;22:5-11
pubmed: 32273666
J Orthop Res. 2014 Mar;32(3):413-22
pubmed: 24249665
Eur J Orthop Surg Traumatol. 2018 Feb;28(2):269-275
pubmed: 29030711
Int Orthop. 2018 Oct;42(10):2329-2334
pubmed: 29468265
Acta Orthop. 2014 Aug;85(4):389-95
pubmed: 24954484
Hip Int. 2010;20 Suppl 7:S63-9
pubmed: 20512775
Hip Int. 2019 May;29(3):270-275
pubmed: 29781288
Hip Int. 2016 Nov 10;26(6):537-542
pubmed: 27739568
Hip Int. 2012 Jul-Aug;22(4):371-8
pubmed: 22865253
J Arthroplasty. 2018 Apr;33(4):1126-1132
pubmed: 29246717
Open Orthop J. 2012;6:578-81
pubmed: 23284594
Clin Orthop Relat Res. 2014 Oct;472(10):3159-65
pubmed: 25030101
Orthop Traumatol Surg Res. 2015 Sep;101(5):571-5
pubmed: 26148967
Orthop Traumatol Surg Res. 2016 Feb;102(1):91-7
pubmed: 26867707
J Orthop Res. 2019 Mar;37(3):681-688
pubmed: 30720205
J Arthroplasty. 2018 Jan;33(1):156-161
pubmed: 28887022
Clin Orthop Relat Res. 2017 Apr;475(4):1045-1054
pubmed: 27752989
Surg Radiol Anat. 2019 Jan;41(1):117-124
pubmed: 30488299
Orthop Traumatol Surg Res. 2010 Jun;96(4):367-75
pubmed: 20471935
EFORT Open Rev. 2019 Nov 1;4(11):626-632
pubmed: 31754469
Int Orthop. 2020 Jan;44(1):45-51
pubmed: 31254017
Acta Orthop Belg. 2010 Aug;76(4):432-42
pubmed: 20973347
Orthop Traumatol Surg Res. 2016 Feb;102(1 Suppl):S9-20
pubmed: 26797005
J Arthroplasty. 1999 Feb;14(2):159-64
pubmed: 10065720
Hip Int. 2007 Jul-Sep;17(3):155-9
pubmed: 19197861
Hip Int. 2011 Jul-Aug;21(4):428-35
pubmed: 21818743
J Orthop Surg Res. 2010 Oct 15;5:76
pubmed: 20950444
J Arthroplasty. 2009 Sep;24(6):990-7
pubmed: 18757175
Int Orthop. 2015 Jul;39(7):1269-75
pubmed: 25522801
Eur J Orthop Surg Traumatol. 2017 Jul;27(5):643-651
pubmed: 28391517

Auteurs

Thierry Scheerlinck (T)

Department of Orthopaedic Surgery and Traumatology, Universitair Ziekenhuis Brussel, Brussels, Belgium.
Medische Beeldvorming en Fysische Wetenschappen (BEFY), Vrije Universiteit Brussel, Brussels, Belgium.

Elien De Winter (E)

Department of Orthopaedic Surgery and Traumatology, Universitair Ziekenhuis Brussel, Brussels, Belgium.
Medische Beeldvorming en Fysische Wetenschappen (BEFY), Vrije Universiteit Brussel, Brussels, Belgium.

Amelie Sas (A)

Department of Mechanical Engineering, Katholieke Universiteit Leuven, Leuven, Belgium.

Sjoerd Kolk (S)

Materialise Group, Leuven, Belgium.

Gert Van Gompel (G)

Medische Beeldvorming en Fysische Wetenschappen (BEFY), Vrije Universiteit Brussel, Brussels, Belgium.
Department of Radiology, Universitair Ziekenhuis Brussel, Brussels, Belgium.

Jef Vandemeulebroucke (J)

Department of Electronics and Informatics (ETRO), Vrije Universiteit Brussel, Brussel, Belgium.
iMinds, Department of Medical IT, Gent, Belgium.

Classifications MeSH